National Instruments Network Card PXI PXITM 1000 User Manual

PXI  
PXI -1000 User Manual  
PXI-1000 User Manual  
January 1998 Edition  
Part Number 321710B-01  
 
 
Important Information  
Warranty  
The PXI-1000 is warranted against defects in materials and workmanship for a period of one year from the date of  
shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace  
equipment that proves to be defective during the warranty period. This warranty includes parts and labor.  
The media on which you receive National Instruments software are warranted not to fail to execute programming  
instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced  
by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not  
execute programming instructions if National Instruments receives notice of such defects during the warranty period.  
National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.  
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside  
of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs  
of returning to the owner parts which are covered by warranty.  
National Instruments believes that the information in this manual is accurate. The document has been carefully reviewed  
for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to  
make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should  
consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages  
arising out of or related to this document or the information contained in it.  
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS  
ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED  
BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE  
CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS,  
OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of the liability of  
National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence.  
Any action against National Instruments must be brought within one year after the cause of action accrues. National  
Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty  
provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow  
the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product;  
owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties,  
or other events outside reasonable control.  
Copyright  
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical,  
including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without  
the prior written consent of National Instruments Corporation.  
Trademarks  
PXIis a trademark of National Instruments Corporation.  
Product and company names listed are trademarks or trade names of their respective companies.  
WARNING REGARDING MEDICAL AND CLINICAL USE OF NATIONAL INSTRUMENTS PRODUCTS  
National Instruments products are not designed with components and testing intended to ensure a level of reliability  
suitable for use in treatment and diagnosis of humans. Applications of National Instruments products involving medical  
or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the part of the  
user or application designer. Any use or application of National Instruments products for or involving medical or clinical  
treatment must be performed by properly trained and qualified medical personnel, and all traditional medical safeguards,  
equipment, and procedures that are appropriate in the particular situation to prevent serious injury or death should always  
continue to be used when National Instruments products are being used. National Instruments products are NOT intended  
to be a substitute for any form of established process, procedure, or equipment used to monitor or safeguard human health  
and safety in medical or clinical treatment.  
 
Compliance  
FCC/DOC Radio Frequency Interference  
Class A Compliance  
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance  
with the instructions in this manual, may cause interference to radio and television reception. Classification  
requirements are the same for the Federal Communications Commission (FCC) and the Canadian  
Department of Communications (DOC). This equipment has been tested and found to comply with the  
following two regulatory agencies:  
Federal Communications Commission  
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant  
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful  
interference when the equipment is operated in a commercial environment. This equipment generates,  
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction  
manual, may cause harmful interference to radio communications. Operation of this equipment in a  
residential area is likely to cause harmful interference in which case the user will be required to correct the  
interference at his own expense.  
Notices to User: Changes or modifications not expressly approved by National Instruments could void  
the user’s authority to operate the equipment under the FCC Rules.  
This device complies with the FCC rules only if used with shielded interface cables  
of suitable quality and construction. National Instruments used such cables to test  
this device and provides them for sale to the user. The use of inferior or nonshielded  
interface cables could void the user’s authority to operate the equipment under the  
FCC rules.  
If necessary, consult National Instruments or an experienced radio/television technician for additional  
suggestions. The following booklet prepared by the FCC may also be helpful: Interference to Home  
Electronic Entertainment Equipment Handbook. This booklet is available from the U.S. Government  
Printing Office, Washington, DC 20402.  
Canadian Department of Communications  
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment  
Regulations.  
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur  
du Canada.  
 
For Your Safety  
Caution  
Before undertaking any troubleshooting, maintenance, or exploratory procedure,  
read carefully the WARNING and CAUTION notices.  
!
capable of inflicting personal injury.  
Mainframe Grounding—The PXI-1000 mainframe requires a  
connection from the premise wire safety ground to the PXI-1000  
chassis ground. The earth safety ground must be connected during use  
of this equipment to minimize shock hazards. Refer to the Connecting  
Safety Ground section of Chapter 2, Installation and Configuration,  
for instructions on connecting safety ground.  
Live Circuits—Operating personnel and service personnel must not  
remove protective covers when operating or servicing the PXI-1000.  
Adjustments and service to internal components must be undertaken  
by qualified service technicians. During service of this product, the  
mains connector to the premise wiring must be disconnected.  
Dangerous voltages may be present under certain conditions; use  
extreme caution.  
Explosive Atmosphere—Do not operate the mainframe in conditions  
where flammable gases are present. Under such conditions this  
equipment is unsafe and may ignite the gases or gas fumes.  
Part Replacement—Only service this equipment with parts that are  
exact replacements, both electrically and mechanically. Contact  
National Instruments for replacement part information. Installation of  
parts with those that are not direct replacements may cause harm to  
personnel operating the mainframe. Furthermore, damage or fire may  
occur if replacement parts are unsuitable.  
Modification—Do not modify any part of the mainframe from its  
original condition. Unsuitable modifications may result in safety  
hazards.  
 
 
About This Manual  
Conventions Used in This Manual.................................................................................xi  
Chapter 1  
Getting Started  
System Controller Slot ....................................................................................1-5  
Star Trigger Slot ..............................................................................................1-5  
Local Bus.........................................................................................................1-6  
Chapter 2  
Connecting Safety Ground.............................................................................................2-3  
Connecting to AC Mains Power and Testing Power up ................................................2-3  
Installing PXI Modules..................................................................................................2-6  
Chapter 3  
Maintenance  
Service Interval..............................................................................................................3-1  
Preparation.....................................................................................................................3-1  
© National Instruments Corporation  
vii  
PXI-1000 User Manual  
 
Contents  
Cleaning......................................................................................................................... 3-1  
Interior Cleaning ............................................................................................. 3-2  
Cleaning the Fan Filters................................................................................................. 3-2  
Resetting the AC Mains Circuit Breaker....................................................................... 3-3  
Troubleshooting the PXI-1000...................................................................................... 3-4  
Appendix A  
Specifications  
Appendix B  
Pinouts  
Customer Communication  
Glossary  
Index  
Figures  
Figure 1-2.  
Rear View of the PXI-1000 Mainframe................................................ 1-5  
Figure 2-1.  
Figure 2-3.  
PXI-1000 Mainframe Airflow Side View............................................. 2-2  
Injector/Ejector Handle Position during Module Insertion................... 2-7  
Tables  
Table 1-1.  
Power Cables ........................................................................................ 1-1  
Table 2-1.  
Table 2-2.  
Power Supply Voltages at Power Monitoring Connector (DB-9) ........ 2-4  
DB-9 Connector Pinout......................................................................... 2-5  
Table 3-1.  
Troubleshooting .................................................................................... 3-4  
PXI-1000 User Manual  
viii  
© National Instruments Corporation  
 
Table A-2.  
DC Output Specifications......................................................................A-2  
Table B-1.  
Table B-2.  
Table B-3.  
Table B-4.  
Table B-5.  
Table B-6.  
P1 (J1) Connector Pinout for the System Controller Slot .....................B-2  
P2 (J2) Connector Pinout for the System Controller Slot .....................B-3  
P1 (J1) Connector Pinout for the Star Trigger Slot...............................B-4  
P2 (J2) Connector Pinout for the Star Trigger Slot...............................B-5  
P1 (J1) Connector Pinout for the Peripheral Slot..................................B-6  
P2 (J2) Connector Pinout for the Peripheral Slot..................................B-7  
© National Instruments Corporation  
ix  
PXI-1000 User Manual  
 
About This Manual  
The PXI-1000 User Manual describes the features of the PXI-1000  
mainframe and contains information about configuring the mainframe,  
installing the modules, and operating and using the PXI-1000.  
Organization of This Manual  
This manual is organized as follows:  
Chapter 1, Getting Started, describes the key features of the PXI-1000,  
lists the contents of your kit, and lists optional equipment you can  
order from National Instruments.  
Chapter 2, Installation and Configuration, describes how to prepare  
and operate your PXI-1000 mainframe.  
Chapter 3, Maintenance, describes basic maintenance procedures you  
can perform on the PXI-1000 mainframe.  
Appendix A, Specifications, contains complete specifications for the  
PXI-1000 mainframe.  
Appendix B, Pinouts, describes the P1 and P2 connector pinouts for  
the PXI-1000 backplane.  
Appendix C, Customer Communication, contains forms you can use to  
request help from National Instruments or to comment on our products  
and manuals.  
The Glossary lists abbreviations, acronyms, metric prefixes,  
mnemonics, symbols, and terms.  
The Index contains an alphabetical list of key terms and topics used in  
this manual, including the page where you can find each one.  
Conventions Used in This Manual  
The following conventions are used in this manual:  
This icon to the left of bold italicized text denotes a note, which alerts you  
to important information.  
This icon to the left of bold italicized text denotes a caution, which advises  
you of precautions to take to avoid injury, data loss, or a system crash.  
!
© National Instruments Corporation  
xi  
PXI-1000 User Manual  
 
     
About This Manual  
This icon to the left of bold italicized text denotes a warning, which advises  
you of precautions to take to avoid being electrically shocked.  
bold italic  
Bold italic text denotes a note, caution, or warning.  
italic  
Italic text denotes emphasis, a cross reference, or an introduction to a key  
concept. This font also denotes text from which you supply the appropriate  
word or value, as in Windows 3.x.  
monospace  
Text in this font denotes text or characters that are to be literally enter from  
the keyboard, sections of code, programming examples, and syntax  
examples. This font is also used for the proper names of disk drives, paths,  
directories, device names, functions, variables, filenames, and extensions.  
Related Documentation  
The following documents contain information that you may find helpful as  
you read this manual:  
Compact PCI Specification PICMG 2.0 R2.1  
PXI Specification Revision 1.0  
IEEE 1101.1-1991, IEEE Standard for Mechanical Core  
Specifications for Microcomputers Using IEC 603-2 Connectors  
IEEE 1101.10 and P1101.11, IEEE Standard for Additional  
Mechanical Specifications for Microcomputers Using IEEE 1101.1  
Equipment Practice  
Customer Communication  
National Instruments wants to receive your comments on our products  
and manuals. We are interested in the applications you develop with our  
products, and we want to help if you have problems with them. To make it  
easy for you to contact us, this manual contains comment and configuration  
forms for you to complete. These forms are in Appendix C, Customer  
Communication, at the end of this manual.  
PXI-1000 User Manual  
xii  
© National Instruments Corporation  
 
   
1
Getting Started  
This chapter describes the key features of the PXI-1000, lists the contents  
of your kit, and lists optional equipment you can order from National  
Instruments.  
Unpacking  
Carefully inspect the shipping container and the mainframe for damage.  
Check for visible damage to the metal work. Check to make sure all  
handles, hardware, and switches are undamaged. Inspect the inner chassis  
for any possible damage, debris, or detached components. If damage  
appears to have been caused in shipment, file a claim with the carrier.  
Retain the packing material for possible inspection and/or reshipment.  
What You Need to Get Started  
The PXI-1000 kit contains the following items:  
PXI-1000 mainframe  
Filler panels  
Power cable (see Table 1-1)  
PXI-1000 User Manual  
Floppy disk with Chassis Initialization file, chassis.ini  
Table 1-1. Power Cables  
Power Cable  
Standard 120 V (USA)  
Switzerland 220 V  
Australia 240 V  
Reference Standards  
ANSI C73.11/NEMA 5-15-P/IEC83  
SEV  
AS C112  
© National Instruments Corporation  
1-1  
PXI-1000 User Manual  
 
             
Chapter 1  
Getting Started  
Table 1-1. Power Cables (Continued)  
Power Cable Reference Standards  
Universal Euro 240 V  
North America 240 V  
United Kingdom 240 V  
CEE (7), II, IV, VII IEC83  
ANSI C73.20/NEMA 5-15-P/IEC83  
BS 1363/IEC83  
If you are missing any of the above items or if you have the incorrect power  
cord, contact National Instruments.  
Optional Equipment  
An optional rack-mount kit is available from National Instruments. You  
can use this kit to install the PXI-1000 mainframe into a standard 19 in.  
(482 mm) instrument cabinet. Contact National Instruments to order the  
PXI-1000 rack-mount kit.  
Key Features  
The PXI-1000 combines a high-performance 8-slot PXI backplane with a  
high-output power supply and a structural design that has been optimized  
for maximum usability in a wide range of applications. The mainframe’s  
modular design ensures the highest level of maintainability resulting in a  
very low mean time to repair (MTTR). The PXI-1000 fully complies with  
the PXI Specification, Revision 1.0, offering advanced timing and  
synchronization features.  
The key features of the PXI-1000 include:  
PXI and CompactPCI (PICMG 2.0 R 2.1) module compatibility  
Compact 3U-sized, 8-slot chassis  
300 W of usable power  
Universal AC input: auto-voltage and auto-frequency ranging  
Over-current protection via push-reset circuit breaker (no loose fuses  
to replace)  
Removable modular power supply with MTBF of 90,000 hours  
Remote power status and inhibit via a rear panel connector  
On/Off (Standby) switch located on the front panel for easy access  
PXI-1000 User Manual  
1-2  
© National Instruments Corporation  
 
   
Chapter 1  
Getting Started  
Selectable fan speed for maximum cooling or quiet operation  
Carrying handle for portability  
Tilt feet for bench-top applications  
PXI-1000 Backplane Overview  
Interoperability with CompactPCI  
The PXI-1000 backplane is interoperable with PXI-compatible products  
and standard CompactPCI products. This is an important feature, as many  
PXI-compatible systems may not require components that do not  
implement PXI-specific features. For example, you may want to use a  
standard CompactPCI network interface card in a PXI chassis.  
The signals on the P1 connector of the backplane meet the requirements of  
the CompactPCI specification for both the peripheral and system modules.  
The PXI-specific signals are located on P2 and are only found on the  
signals that are reserved or not used in the CompactPCI 64-bit  
specification. Therefore, all modules that meet the requirements of the  
CompactPCI 64-bit specification will function in the PXI-1000.  
Figures 1-1 and 1-2 show some of the key features and components of the  
PXI-1000 mainframe. Figure 1-1 shows the front view of the PXI-1000 and  
Figure 1-2 shows the rear view.  
© National Instruments Corporation  
1-3  
PXI-1000 User Manual  
 
   
Chapter 1  
Getting Started  
1
ON  
STANDBY  
22  
P2  
1
25  
2
P1  
1
ZABCDEF  
2
3
4
5
6
7
8
1
5
3
4
8
7
6
4
3
1
2
Remote On/Off (Standby) Switch 3 Optional Mounting Brackets  
6
7
8
Star Trigger Slot  
Backplane Connectors  
(Located in Slots 1-8)  
4
5
Removable Feet  
Peripheral Slots  
System Controller Slot  
Controller Expansion Slots  
Figure 1-1. Front View of the PXI-1000 Mainframe  
PXI-1000 User Manual  
1-4  
© National Instruments Corporation  
 
   
Chapter 1  
Getting Started  
1
2
3
90-264 VAC  
47-63 HZ  
8A MAX  
CB1  
10A  
DISCONNECT  
POWEORD  
BEFORE REMOVING  
POWER MODULE  
HI  
LO  
FAN SPEED  
DB-9  
7
6
5
4
1
2
Module Cooling Air Outlets  
Universal AC Input  
3
4
Main Power Switch  
Circuit Breaker  
5
Chassis Ground Screw  
7
Fan Speed  
Selector  
6
Remote Power Status and  
Inhibit Connector  
Figure 1-2. Rear View of the PXI-1000 Mainframe  
System Controller Slot  
The System Controller slot is located in Slot 1 of the chassis as defined by  
the PXI specification. It has three controller expansion slots, which are  
used for system controller modules that are wider than one slot. As defined  
in the PXI specification, these slots allow the controller to expand to the left  
to prevent the controller from using up peripheral slots.  
Star Trigger Slot  
The Star Trigger (ST) slot is located at Slot 2. This slot has a dedicated  
trigger line between each peripheral slot (see Figure 1-3). This slot is  
intended for modules with ST functionality that can provide individual  
triggers to all other peripherals. However, if you do not require advanced  
trigger functionality, you can install any standard peripheral module into  
this slot.  
Peripheral Slots  
There are seven peripheral slots including the Star Trigger controller slot.  
© National Instruments Corporation  
1-5  
PXI-1000 User Manual  
 
         
Chapter 1  
Getting Started  
Local Bus  
The PXI backplane’s local bus is a daisy-chained bus that connects each  
peripheral slot with its adjacent peripheral slots to the left and right, as  
shown in Figure 1-3.  
For example, a given peripheral slot’s right local bus connects to the  
adjacent slot’s left local bus and so on. Each local bus is 13 lines wide and  
can pass analog signals between cards or provide a high-speed side-band  
communication path that does not affect the PXI bandwidth.  
Local Bus signals may range from high-speed TTL signals to analog  
signals as high as 42 V. Initialization software keys adjacent boards to  
prohibit the use of incompatible boards. This software uses the  
configuration information specific to each peripheral board to evaluate  
compatibility. This method is a flexible way to define local bus  
functionality that is not limited by hardware keying.  
Star Triggers  
Local  
Bus  
Local  
Bus  
Local  
Bus  
PCI Arbitration and Clock Signals  
Figure 1-3. PXI Local Bus and Star Trigger Routing  
Trigger Bus  
The eight PXI trigger lines are bused to each slot. You can use the trigger  
lines in a variety of ways. For example, you can use triggers to synchronize  
the operation of several different PXI peripheral modules. In other  
applications, one module can control carefully timed sequences of  
PXI-1000 User Manual  
1-6  
© National Instruments Corporation  
 
       
Chapter 1  
Getting Started  
operations performed on other modules in the system. Modules can pass  
triggers to one another, allowing precisely timed responses to  
asynchronous external events the system is monitoring or controlling.  
System Reference Clock  
The PXI-1000 supplies the PXI 10 MHz system clock signal (PXI_CLK10)  
independently to each peripheral slot. An independent buffer (having a  
source impedance matched to the backplane and a skew of less than 1 ns  
between slots) drives the clock signal to each peripheral slot. You can use  
this common reference clock signal to synchronize multiple modules in a  
measurement or control system. You can drive PXI_CLK10 from an  
external source through the PXI_CLK10_IN pin on the P2 connector of the  
Star Trigger Slot. (See Table B-1, P1 (J1) Connector Pinout for the System  
Controller Slot, in Appendix B, Pinouts.) Sourcing an external clock on  
this pin automatically disables the backplane’s 10 MHz source.  
© National Instruments Corporation  
1-7  
PXI-1000 User Manual  
 
 
2
Installation and Configuration  
This chapter describes how to prepare and operate your PXI-1000  
mainframe.  
Before connecting the mainframe to a power source, read this chapter and  
the For Your Safety section located at the beginning of this manual.  
Site Considerations  
The PXI-1000 is designed to operate on a bench or in an instrument rack.  
Determine how you want to use your PXI-1000 and follow the appropriate  
installation instructions.  
Apertures in the rear and along both sides of the mainframe facilitate power  
supply and module cooling. Air enters through filters and fan inlets located  
in the lower rear of the mainframe and exits through the upper sections on  
both sides and through the rear, as shown in Figure 2-1. Place your  
PXI-1000 on a bench top or in an instrument rack so that the fans (air inlets)  
and the air outlet apertures along both sides of the mainframe have adequate  
ventilation. Keep other equipment a minimum of 3.0 in. (76.2 mm) away  
from the air inlets and outlets.  
© National Instruments Corporation  
2-1  
PXI-1000 User Manual  
 
       
Chapter 2  
Installation and Configuration  
1
2
1
Module Cooling Air Outlets  
2
Module Cooling Air Intake  
Figure 2-1. PXI-1000 Mainframe Airflow Side View  
Install your mainframe so that you can easily access the rear panel. This  
simplifies the replacement of the air filters or power supply/fan assembly,  
if necessary.  
Rack Mounting  
Rack-mount applications require the optional rack-mount kit available  
from National Instruments. Refer to the instructions supplied with the  
rack-mount kit to install your PXI-1000 in an instrument rack.  
Note  
You may wish to remove the feet from your PXI-1000 when rack mounting. To do  
so, remove the two screws holding the feet in place.  
Setting Fan Speed  
The fan speed selector switch is located on the rear panel of the PXI-1000.  
Refer to Figure 1-2, Rear View of the PXI-1000 Mainframe, for a diagram  
of the fan speed selector. Select HI for maximum cooling effectiveness  
(recommended) or LO for more quiet operation.  
PXI-1000 User Manual  
2-2  
© National Instruments Corporation  
 
       
Chapter 2  
Installation and Configuration  
Connecting Safety Ground  
Warning The PXI-1000 chassis is designed with a three-position NEMA 15-5 style plug that  
connects the ground line to the chassis ground. To minimize shock hazard, make  
sure your electrical power outlet has an appropriate earth safety ground that is  
connected whenever you power up the chassis.  
If your power outlet does not have an appropriate ground connection, you  
must connect the premise wire safety ground to the chassis grounding  
screw located on the rear panel. Refer to Figure 1-2, Rear View of the  
PXI-1000 Mainframe, for a diagram of the chassis grounding screw. To  
connect the safety ground, complete the following steps:  
1. Connect a 16 AWG (1.3 mm) wire to the chassis grounding screw  
using a toothed grounding lug. The wire must have green insulation  
with a yellow stripe or must be non-insulated (bare).  
2. Attach the opposite end of the wire to permanent earth ground using  
toothed washers or a toothed lug.  
Connecting to AC Mains Power and Testing Power up  
Caution  
Make sure the main power switch located on the rear panel of the PXI-1000 is in  
the Off position (0). Refer to Figure 1-2, Rear View of the PXI-1000 Mainframe, for  
a diagram of the main power switch.  
!
Do not install modules prior to performing the first power-on test.  
The power supply is universal, which means the mainframe can connect to  
all standard worldwide input voltages. Refer to Chapter 1, Getting Started,  
for power cord specifications. Attach input power through the rear AC inlet  
using the appropriate line cord supplied. Refer to Figure 1-2, Rear View of  
the PXI-1000 Mainframe, for a diagram of the IEC 320 inlet.  
Push the main power switch to the On (1) position. Then push the remote  
power switch to the On position (if not already on). Observe that all fans  
become operational. Note that the main power switch (rear panel) enables  
the power module to receive AC power. The remote front panel standby  
switch allows the power module to provide power to the PXI backplane  
(if the rear main power switch is on). For this reason, you can leave the  
main power switch on and use the remote On/Off (Standby) switch on the  
front to power up and power down the backplane.  
© National Instruments Corporation  
2-3  
PXI-1000 User Manual  
 
       
Chapter 2  
Installation and Configuration  
Caution  
When connecting digital voltmeter probes to the rear D-sub connector, be careful  
not to short the probe leads together. Doing so could damage the power supply.  
!
You can use a digital voltmeter to ensure all voltage levels in your  
PXI-1000 are within the allowable limits. Referring to Table 2-1, connect  
one lead of the voltmeter to a supply pin on the remote power monitoring  
connector (9-pin D-sub) located on the rear panel. Refer to Table 2-2 for a  
pinout diagram of the remote power monitoring connector. Connect the  
reference lead of the voltmeter to one of the ground pins. Compare each  
voltage reading to the values listed in Table 2-1.  
Note  
Use the rear panel D-sub connector to check voltages only. Do not use the  
connector to supply power to external devices.  
Table 2-1. Power Supply Voltages at Power Monitoring Connector (DB-9)  
Pin  
2
Supply  
+5 V  
Acceptable Voltage Range  
4.75 to 5.25 V  
3.135 to 3.465 V  
11.4 to 12.6 V  
–12.6 to –11.4 V  
N/A  
4
+3.3 V  
6
+12 V  
8
–12 V  
1, 9  
Logic Ground  
If the voltages fall within the specified ranges, the mainframe complies  
with the CompactPCI voltage limit specifications. Note that the rear panel  
D-sub connector is to be used to check voltages only. Do not use these  
voltages to supply power to external devices.  
Note  
If the fans or power unit fail to function properly, refer to the Troubleshooting the  
PXI-1000 section of Chapter 3, Maintenance.  
PXI-1000 User Manual  
2-4  
© National Instruments Corporation  
 
   
Chapter 2  
Installation and Configuration  
Remote Power Monitoring and Inhibiting Interface  
The PXI-1000 mainframe supports remote power monitoring and  
inhibiting via a 9-pin D-sub connector located on the rear panel. Table 2-2  
shows the pinout of the DB-9 connector.  
Table 2-2. DB-9 Connector Pinout  
DB-9 Pin  
Signal  
1
2
3
4
5
6
7
8
9
Logic Ground  
+5 V  
Reserved  
+3.3 V  
Inhibit*  
+12 V  
Reserved  
–12 V  
Logic Ground  
5
4
3
2
1
9
8
7
6
You can use the Inhibit signal (active low) to turn off the power supply  
outputs. To use this feature, connect the Inhibit pin (pin 5) to a Logic  
Ground pin (pin 1 or 9). Make sure both the rear main power switch and  
the front (standby) switch are in the ON position. As long as Inhibit is  
pulled down, the power supply inhibits its DC outputs. DC output resumes  
when Inhibit is no longer connected to ground. Note that the power  
(standby) switch, located on the front of the chassis, uses this inhibiting  
feature. For remote reset, connect a momentary switch between pin 5 and  
pin 1 (or pin 9).  
© National Instruments Corporation  
2-5  
PXI-1000 User Manual  
 
     
Chapter 2  
Installation and Configuration  
Installing PXI Modules  
Caution  
Turn off the mainframe power before installing CompactPCI or PXI modules.  
!
Install a module into a mainframe slot by first placing the module’s card  
edges into the front module guides (top and bottom), as shown in  
Figure 2-2. Slide the module to the rear of the mainframe (making sure that  
the injector/ejector handle is pushed down as shown in Figure 2-3). When  
you begin to feel resistance, push up on the injector/ejector handle to inject  
the card into the frame. Secure the module’s front panel to the mainframe  
using the module’s front-panel mounting screws.  
PXI-1000 Chassis  
PXI-8156  
System Controller  
ON  
STANDBY  
1
2
3
4
5
6
7
8
PXI Board  
Injector/Ejector Rail  
Injector/Ejector Handle  
Figure 2-2. Installing PXI or CompactPCI Modules  
PXI-1000 User Manual  
2-6  
© National Instruments Corporation  
 
     
Chapter 2  
Installation and Configuration  
Figure 2-3. Injector/Ejector Handle Position during Module Insertion  
Installing Filler Panels  
To optimize module cooling performance, install filler panels into unused  
or empty slots. Secure with the captive mounting screws.  
Using the Chassis Initialization File  
To assist system integrators, the PXI specification requires manufacturers  
of PXI chassis and system modules to document the capabilities of their  
products. The minimum documentation requirements are contained in  
.ini files, which consist of ASCII text. The system integrator can read the  
.ini file, and configuration utilities and device drivers can also use this  
file. The PXI-1000 chassis initialization file, chassis.ini, is included on  
the diskette for your PXI-1000.  
© National Instruments Corporation  
2-7  
PXI-1000 User Manual  
 
       
3
Maintenance  
This chapter describes basic maintenance procedures you can perform on  
the PXI-1000 mainframe.  
Service Interval  
Clean the mainframe fan filters at a maximum interval of six months.  
Depending upon the amount of use and ambient dust levels in the operating  
environment, the filters may require more frequent cleaning.  
Clean dust from the mainframe exterior (and interior) as needed, based on  
the operating environment. Periodic cleaning increases reliability.  
Preparation  
The information in this section is designed for use by qualified service  
personnel. Read the For Your Safety section at the beginning of this manual  
before attempting any procedures in this chapter.  
Caution  
Many components within the mainframe are susceptible to static discharge  
damage. Service the mainframe only in a static-free environment. Observe  
standard handling precautions for static-sensitive devices while servicing the  
mainframe. Always wear a grounded wrist strap, or equivalent, while servicing  
the mainframe.  
!
Cleaning  
Cleaning procedures consist of exterior and interior cleaning of the  
mainframe and cleaning the fan filters. Refer to your module user  
documentation for information on cleaning the individual CompactPCI  
or PXI modules.  
Caution  
Always power-off the mainframe (using the main power switch on the rear of the  
mainframe) and disconnect the power cord before cleaning or servicing the  
mainframe.  
!
© National Instruments Corporation  
3-1  
PXI-1000 User Manual  
 
           
Chapter 3  
Maintenance  
Interior Cleaning  
Exterior Cleaning  
Use a dry, low-velocity stream of air to clean the interior of the mainframe.  
Use a soft-bristle brush for cleaning around components. If you must use a  
liquid for minor interior cleaning, use a 75% isopropyl alcohol solution and  
rinse with deionized water.  
Clean the exterior surfaces of the mainframe with a dry lint-free cloth or a  
soft-bristle brush. If any dirt remains, wipe with a cloth moistened in a mild  
soap solution. Remove any soap residue by wiping with a cloth moistened  
with clear water. Do not use abrasive compounds on any part of the  
mainframe.  
Cautions Avoid getting moisture inside the mainframe during exterior cleaning. Use just  
!
enough moisture to dampen the cloth.  
Do not wash the front- or rear-panel connectors or switches. Cover these  
components while cleaning the mainframe.  
Do not use chemical cleaning agents; they may damage the mainframe. Avoid  
chemicals that contain benzene, toluene, xylene, acetone, or similar solvents.  
Cleaning the Fan Filters  
You can easily remove the mainframe cooling filters from the rear of the  
mainframe by removing the plastic housing attached to each fan.  
Clean the fan filters by washing them in a mild soap solution and then  
vacuuming or blowing air through them. Rinse the filters with water and  
allow them to dry before reinstalling them on the mainframe.  
PXI-1000 User Manual  
3-2  
© National Instruments Corporation  
 
     
Chapter 3  
Maintenance  
Resetting the AC Mains Circuit Breaker  
If your PXI-1000 encounters an over-current condition, the circuit breaker  
located on the rear panel will trip to prevent damage to the mainframe.  
Complete the following steps to reset the circuit breaker:  
1. Turn the front panel (remote) power switch to the Standby position.  
2. Turn off the rear main power switch.  
3. Depress the circuit breaker to reset it.  
4. Turn on the rear main power switch.  
5. Turn the front panel (remote) power switch to the On position.  
If the circuit breaker trips again, complete the following steps:  
1. Turn the front panel power switch to the Standby position.  
2. Turn off the rear main power switch.  
3. Disconnect the mainframe from the AC mains power source.  
4. Remove all modules from the mainframe.  
5. Complete the test procedure described in the Connecting to AC Mains  
Power and Testing Power up section in Chapter 2, Installation and  
Configuration.  
6. If any voltages are outside the acceptable limits, contact National  
Instruments.  
7. If all voltages are within the acceptable limits, verify that your  
PXI-1000 can meet the power requirements of your CompactPCI or  
PXI modules. Overloading the chassis can cause the breaker to trip.  
Refer to Appendix A, Specifications.  
8. The over-current condition that caused the circuit breaker to trip may  
be due to a faulty CompactPCI or PXI module. Refer to the  
documentation that was supplied with the modules for troubleshooting  
your modules.  
© National Instruments Corporation  
3-3  
PXI-1000 User Manual  
 
 
Chapter 3  
Maintenance  
Troubleshooting the PXI-1000  
Refer to Table 3-1 to troubleshoot the PXI-1000 mainframe. The table lists  
possible causes for power failure and recommends ways to correct the  
problem.  
Table 3-1. Troubleshooting  
Possible Cause  
What to Do  
PXI-1000 mainframeis not Make sure that the PXI-1000 is connected  
connected to power source. to a live electrical outlet. Try operating  
another piece of equipment from this  
outlet.  
Power switches are not  
switched on.  
Set the rear (main) power switch and the  
front (standby) power switch to the On  
position.  
Remote inhibiting input on Deactivate your system’s remote  
the rear panel of the  
mainframe is active.  
inhibiting control.  
Circuit breaker is tripped.  
Reset the circuit breaker. Refer to the  
Resetting the AC Mains Circuit Breaker  
section in this chapter.  
Power supply has failed.  
Contact National Instruments.  
PXI-1000 User Manual  
3-4  
© National Instruments Corporation  
 
       
A
Specifications  
This appendix contains complete specifications for the PXI-1000  
mainframe.  
Electrical  
Table A-1. AC Input Specifications  
Characteristic Description  
Input Voltage Range  
Input Frequency Range  
Over-Current Protection  
Operating Current  
90–264 VAC  
47 to 63 Hz  
10 A circuit breaker  
8 A  
Line Regulation  
± 0.5% over operating line range, all  
outputs  
Efficiency  
80% typical  
Transient Response  
Output maximum excursion of ±5%  
for 25% load step. Recovery less than  
500 µs.  
Power Disconnect  
The rear main power switch supplies AC  
power to power module. The front  
(standby) power switch causes the power  
module to supply DC power to the  
CompactPCI/PXI backplane whenever  
the main power switch (rear panel) is in  
the On (1) position. The rear-panel D-sub  
connector facilitates remote inhibiting  
operation. Both the rear main power  
switch and the front (standby) switch  
must be in the On position prior to use of  
remote inhibit. The power cord provides  
main power disconnect.  
© National Instruments Corporation  
A-1  
PXI-1000 User Manual  
 
       
Appendix A  
Specifications  
Table A-2. DC Output Specifications  
Characteristic Description  
Maximum Usable Power  
DC Current Capacity (IMP  
300 W  
Voltage  
+3.3 V  
+12 V  
+5 V  
)
IMP (Steady-State Current)  
40 A  
4 A  
20 A  
1 A  
–12 V  
Load Regulation  
1% for 10% to 100% load changes,  
all outputs  
10% min. load required on V1 (+3.3 V)  
for max. load regulation on outputs  
V2 (+5 V), V3 (+12 V), and V4 (–12 V).  
Maximum Ripple and  
Noise  
1% ripple, 1% noise, or 100 mV,  
whichever is greater  
20 MHz bandwidth  
Over-Current Protection  
Over-Voltage Protection  
All outputs protected from short circuit  
and overload, automatic recovery  
3.3 V clamps at 20% to 30% above output  
voltage  
5 V, +12 V, and –12 V clamps at 10% to  
20% above output voltage  
Power Supply Unit MTBF 90,000+ hr  
Power Supply/Fan Unit  
MTTR  
Replacement in under 5 minutes  
PXI-1000 User Manual  
A-2  
© National Instruments Corporation  
 
 
Appendix A  
Specifications  
Cooling  
Table A-3. Cooling Specifications  
Characteristic Description  
Per Slot Cooling Capacity Slot cooling capacity in worst-case slot  
is 20 W with fan speed set to HI  
Module Cooling System  
Forced air circulation (positive  
pressurization) via two 60 cfm fans with  
HI/LO speed selector  
Slot Airflow Direction  
P1 to P2, bottom of module to top of  
module  
Module Cooling Intake  
Module Cooling Exhaust  
Bottom rear of mainframe  
Along both sides of mainframe  
Forced air circulation via integrated fan  
Power Supply Cooling  
System  
Power Supply Cooling  
Intake  
Rear of mainframe  
Power Supply Cooling  
Exhaust  
Along both sides of mainframe rear and  
upper rear panel  
Module Cooling Fan  
MTBF  
40,000+ hr  
Power Supply/Fan Unit  
Replacement in under 5 minutes  
Safety  
Table A-4. Safety Specifications  
Characteristic  
Description  
Safety Characteristics  
UL 3111-1, IEC 1010-1, CSA 22.2  
No. 1010.1  
Installation Category II  
Pollution Degree 2  
Safety Class 1  
© National Instruments Corporation  
A-3  
PXI-1000 User Manual  
 
 
Appendix A  
Specifications  
Environmental  
Table A-5. Environmental Specifications  
Characteristic  
Description  
Operating Temperature  
0° to 50° C (fan in LO position)  
0° to 55° C (fan in HI position)  
Storage Temperature  
–20° to 70° C  
Operating Relative  
Humidity  
Maximum 80% for temperatures up to  
31° C, decreasing linearly to 50% at 40° C  
Functional Shock  
(Operating)  
MIL-T-28800E CLASS 3, Half-Sine  
Shock Pulse, 11 ms duration, 30 g peak  
Operating Location  
Indoor use  
Random Vibration  
(Operational)*  
5 to 500 Hz, 0.31 gRMS  
Random Vibration  
(Non-Operational)*  
10 to 500 Hz, 2.46 gRMS  
EMC Emissions  
EMC Immunity  
Altitude  
FCC Class A compliant and EN 55011  
Group 1 Class A Compliant  
Refer to DOC supplied with chassis for  
compliance to relevant directives.  
2 km (1.24 mi)  
* Random vibration profiles were developed in accordance with MIL-T-28800E CLASS 3  
and MIL-STD-810E Method 514 Test levels exceed those recommended in  
MIL-STD-810E for Category 1 (Basic Transportation), Figures 514.4-1 through 514.4-3.  
PXI-1000 User Manual  
A-4  
© National Instruments Corporation  
 
 
Appendix A  
Specifications  
Backplane  
Table A-6. Backplane Specifications  
Characteristic Description  
Size  
3U-sized; one system slot (with three  
system expansion slots) and seven  
peripheral slots.  
Compliant with IEEE 1101.10  
mechanical packaging.  
PXI Specification Revision 1.0  
compliant.  
Accepts both PXI and CompactPCI  
(PICMG 2.0 R2.1) 3U modules.  
Backplane Bare-Board  
Material  
UL 94 V-0 recognized  
(File No. E 116551)  
Backplane Connectors  
Conform to IEC 917 and IEC 1076-4-101,  
and are UL 94 V-0 rated  
© National Instruments Corporation  
A-5  
PXI-1000 User Manual  
 
 
Appendix A  
Specifications  
Mechanical  
Table A-7. Mechanical Specifications  
Characteristic Description  
Overall Dimensions  
Standard Mainframe  
Height  
Width  
Depth  
17.78 cm (7.00 in.)  
27.02 cm (10.64 in.)  
37.85 cm (14.90 in.)  
Notes:  
1.80 cm (.71 in.) added to height when  
feet are installed.  
When tilted with front feet extended on  
table top, height is increased  
approximately 5.29 cm (2.083 in.) in front  
and 1.48 cm (.583 in.) in rear.  
Weight  
8.6 kg (19 lb.)  
Maximum Module Weight 1.8 kg (4 lb.)  
Materials  
Sheet Aluminum (5052-H32) and  
Cold Rolled Steel  
Finish  
Unpainted Aluminum  
Conductive Clear Iridite  
Clear Chromate Zinc Plating  
Polyurethane Enamel  
Cold Rolled Steel  
Paint  
PXI-1000 User Manual  
A-6  
© National Instruments Corporation  
 
 
Appendix A  
Specifications  
Figure A-1 shows the PXI-1000 dimensions. The holes shown are for the  
installation of the optional rack-mount kit. You can install this kit on the  
front or rear of the chassis, depending on which end of the chassis you wish  
to face toward the front of the instrument cabinet. Note that the front and  
rear chassis mounting holes (size M4) are symmetrical.  
ON  
STANDBY  
7.00 in.  
2
3
4
5
6
7
8
1
10.64 in.  
14.90 in.  
1.823 in.  
.78 in.  
1.00 in.  
3.541 in. 1.844 in.  
.71 in.  
Figure A-1. PXI-1000 Dimensions  
© National Instruments Corporation  
A-7  
PXI-1000 User Manual  
 
   
B
Pinouts  
This appendix describes the P1 and P2 connector pinouts for the PXI-1000  
backplane.  
Table B-1 shows the P1 (J1) connector pinout for the System Controller  
slot.  
Table B-2 shows the P2 (J2) connector pinout for the System Controller  
slot.  
Table B-3 shows the P1 (J1) connector pinout for the Star Trigger slot.  
Table B-4 shows the P2 (J2) connector pinout for the Star Trigger slot.  
Table B-5 shows the P1 (J1) connector pinout for the peripheral slots.  
Table B-6 shows the P2 (J2) connector pinout for the peripheral slots.  
Note  
PXI signals are shown in boldface.  
© National Instruments Corporation  
B-1  
PXI-1000 User Manual  
 
     
Appendix B  
Pinouts  
Table B-1. P1 (J1) Connector Pinout for the System Controller Slot  
Pin  
25  
24  
23  
22  
21  
20  
19  
18  
17  
16  
15  
12–14  
11  
10  
9
Z
A
B
REQ64#  
5V  
C
ENUM#  
V(I/O)  
AD[3]  
3.3V  
D
E
F
GND  
5V  
3.3V  
5V  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[1]  
3.3V  
AD[0]  
5V  
ACK64#  
AD[2]  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[4]  
GND  
AD[7]  
3.3V  
AD[6]  
M66EN  
AD[11]  
GND  
AD[5]  
AD[9]  
GND  
AD[8]  
V(I/O)  
AD[14]  
3.3V  
C/BE[0]#  
AD[10]  
AD[13]  
C/BE[1]#  
PERR#  
LOCK#  
TRDY#  
AD[12]  
3.3V  
AD[15]  
GND  
SERR#  
3.3V  
PAR  
SDONE  
GND  
SBO#  
GND  
DEVSEL#  
3.3V  
V(I/O)  
IRDY#  
Key Area  
AD[16]  
3.3V  
STOP#  
GND  
FRAME#  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[18]  
AD[21]  
C/BE[3]#  
AD[26]  
AD[30]  
REQ#  
AD[17]  
GND  
GND  
AD[20]  
GND  
AD[25]  
GND  
CLK  
C/BE[2]#  
AD[19]  
AD[22]  
AD[24]  
AD[27]  
AD[31]  
GNT#  
INTS  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
IDSEL  
GND  
AD[23]  
V(I/O)  
AD[28]  
3.3V  
8
7
AD[29]  
GND  
6
5
BRSVP1A5  
BRSVP1A4  
INTA#  
BRSVP1B5  
GND  
RST#  
GND  
INTP  
5V  
4
V(I/O)  
INTC#  
TMS  
3
INTB#  
5V  
INTD#  
TDI  
2
TCK  
TDO  
+12V  
1
5V  
–12V  
TRST#  
5V  
PXI-1000 User Manual  
B-2  
© National Instruments Corporation  
 
   
Appendix B  
Pinouts  
Table B-2. P2 (J2) Connector Pinout for the System Controller Slot  
Pin  
22  
21  
20  
19  
18  
17  
16  
15  
14  
13  
12  
11  
10  
9
Z
A
PXI_RSVA22  
RSV  
B
PXI_RSVB22  
GND  
C
PXI_RSVC22  
RSV  
D
PXI_RSVD22  
RSV  
E
PXI_RSVE22  
RSV  
F
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
RSV  
RSV  
RSV  
GND  
RSV  
RSV  
GND  
RSV  
RSV  
RSV  
PXI_TRIG3  
PXI_TRIG2  
PXI_TRIG1  
PXI_BRSVA15  
AD[35]  
PXI_TRIG4  
GND  
PXI_TRIG5  
PRST#  
DEG#  
GND  
PXI_TRIG6  
GNT6#  
PXI_TRIG7  
GNT5#  
AD[32]  
AD[36]  
AD[39]  
AD[43]  
AD[46]  
AD[50]  
AD[53]  
AD[57]  
AD[60]  
PAR64  
REQ6#  
GND  
PXI_TRIG0  
GND  
FAL#  
REQ5#  
GND  
AD[34]  
GND  
AD[33]  
V(I/O)  
AD[38]  
AD[37]  
GND  
AD[42]  
AD[41]  
GND  
AD[40]  
V(I/O)  
AD[45]  
AD[44]  
GND  
AD[49]  
AD[48]  
GND  
AD[47]  
V(I/O)  
AD[52]  
AD[51]  
GND  
8
AD[56]  
AD[55]  
GND  
AD[54]  
V(I/O)  
7
AD[59]  
AD[58]  
GND  
6
AD[63]  
AD[62]  
GND  
AD[61]  
V(I/O)  
5
C/BE[5]#  
V(I/O)  
C/BE[4]#  
GND  
4
PXI_BRSVB4  
GND  
C/BE[7]#  
GNT3#  
SYSEN#  
REQ1#  
C/BE[6]#  
GNT4#  
REQ3#  
REQ2#  
3
CLK4  
REQ4#  
GNT2#  
GNT1#  
2
CLK2  
CLK3  
1
CLK1  
GND  
© National Instruments Corporation  
B-3  
PXI-1000 User Manual  
 
   
Appendix B  
Pinouts  
Table B-3. P1 (J1) Connector Pinout for the Star Trigger Slot  
Pin  
25  
24  
23  
22  
21  
20  
19  
18  
17  
16  
15  
12–14  
11  
10  
9
Z
A
B
REQ64#  
5V  
C
ENUM#  
V(I/O)  
AD[3]  
3.3V  
D
E
F
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
5V  
3.3V  
AD[0]  
5V  
5V  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[1]  
3.3V  
AD[7]  
3.3V  
ACK64#  
AD[2]  
AD[4]  
GND  
AD[6]  
AD[5]  
AD[9]  
GND  
AD[8]  
V(I/O)  
AD[14]  
3.3V  
M66EN  
AD[11]  
GND  
C/BE[0]#  
AD[10]  
AD[13]  
C/BE[1]#  
PERR#  
LOCK#  
TRDY#  
AD[12]  
3.3V  
AD[15]  
GND  
SERR#  
3.3V  
PAR  
SDONE  
GND  
SBO#  
GND  
DEVSEL#  
3.3V  
V(I/O)  
IRDY#  
STOP#  
GND  
FRAME#  
Key Area  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[18]  
AD[21]  
C/BE[3]#  
AD[26]  
AD[30]  
REQ#  
AD[17]  
GND  
AD[16]  
3.3V  
GND  
AD[20]  
GND  
AD[25]  
GND  
CLK  
C/BE[2]#  
AD[19]  
AD[22]  
AD[24]  
AD[27]  
AD[31]  
GNT#  
INTS  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
IDSEL  
GND  
AD[23]  
V(I/O)  
AD[28]  
3.3V  
8
7
AD[29]  
GND  
6
5
BRSVP1A5  
BRSVP1A4  
INTA#  
BRSVP1B5  
GND  
RST#  
GND  
INTP  
5V  
4
V(I/O)  
INTC#  
TMS  
3
INTB#  
5V  
INTD#  
TDI  
2
TCK  
TDO  
+12V  
1
5V  
–12V  
TRST#  
5V  
PXI-1000 User Manual  
B-4  
© National Instruments Corporation  
 
   
Appendix B  
Pinouts  
Table B-4. P2 (J2) Connector Pinout for the Star Trigger Slot  
Pin  
22  
21  
20  
19  
18  
17  
16  
15  
14  
13  
12  
11  
10  
9
Z
A
B
PXI_RSVB22  
GND  
C
PXI_RSVC22  
PXI_LBR1  
PXI_STAR0  
PXI_STAR3  
PXI_TRIG5  
PRST#  
D
PXI_RSVD22  
PXI_LBR2  
GND  
E
F
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
PXI_RSVA22  
PXI_LBR0  
PXI_LBR4  
PXI_STAR2  
PXI_TRIG3  
PXI_TRIG2  
PXI_TRIG1  
PXI_BRSVA15  
AD[35]  
PXI_RSVE22  
PXI_LBR3  
PXI_STAR1  
PXI_STAR5  
PXI_TRIG6  
PXI_CLK10  
PXI_TRIG7  
PXI_LBR6  
AD[32]  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
PXI_LBR5  
GND  
PXI_STAR4  
GND  
PXI_TRIG4  
GND  
PXI_CLK10_IN  
GND  
PXI_TRIG0  
GND  
DEG#  
FAL#  
PXI_STAR6  
GND  
AD[34]  
GND  
AD[33]  
AD[38]  
V(I/O)  
AD[37]  
AD[36]  
AD[42]  
AD[41]  
GND  
AD[40]  
GND  
AD[39]  
AD[45]  
V(I/O)  
AD[44]  
AD[43]  
AD[49]  
AD[48]  
GND  
AD[47]  
GND  
AD[46]  
AD[52]  
V(I/O)  
AD[51]  
AD[50]  
8
AD[56]  
AD[55]  
GND  
AD[54]  
GND  
AD[53]  
7
AD[59]  
V(I/O)  
AD[58]  
AD[57]  
6
AD[63]  
AD[62]  
GND  
AD[61]  
GND  
AD[60]  
5
C/BE[5]#  
V(I/O)  
V(I/O)  
C/BE[4]#  
GND  
PAR64  
4
PXI_BRSVB4  
GND  
C/BE[7]#  
PXI_LBR8  
SYSEN#  
PXI_STAR10  
C/BE[6]#  
PXI_LBR10  
PXI_STAR8  
PXI_STAR12  
3
PXI_LBR7  
PXI_LBR11  
PXI_STAR9  
PXI_LBR9  
PXI_STAR7  
PXI_STAR11  
2
PXI_LBR12  
GND  
1
© National Instruments Corporation  
B-5  
PXI-1000 User Manual  
 
   
Appendix B  
Pinouts  
Table B-5. P1 (J1) Connector Pinout for the Peripheral Slot  
Pin  
25  
24  
23  
22  
21  
20  
19  
18  
17  
16  
15  
12–14  
11  
10  
9
Z
A
B
REQ64#  
5V  
C
ENUM#  
V(I/O)  
AD[3]  
3.3V  
D
E
F
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
5V  
3.3V  
AD[0]  
5V  
5V  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[1]  
3.3V  
AD[7]  
3.3V  
ACK64#  
AD[2]  
AD[4]  
GND  
AD[6]  
AD[5]  
AD[9]  
GND  
AD[8]  
V(I/O)  
AD[14]  
3.3V  
M66EN  
AD[11]  
GND  
C/BE[0]#  
AD[10]  
AD[13]  
C/BE[1]#  
PERR#  
LOCK#  
TRDY#  
AD[12]  
3.3V  
AD[15]  
GND  
SERR#  
3.3V  
PAR  
SDONE  
GND  
SBO#  
GND  
DEVSEL#  
3.3V  
V(I/O)  
IRDY#  
STOP#  
GND  
FRAME#  
Key Area  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
AD[18]  
AD[21]  
C/BE[3]#  
AD[26]  
AD[30]  
REQ#  
AD[17]  
GND  
AD[16]  
3.3V  
GND  
AD[20]  
GND  
AD[25]  
GND  
CLK  
C/BE[2]#  
AD[19]  
AD[22]  
AD[24]  
AD[27]  
AD[31]  
GNT#  
INTS  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
IDSEL  
GND  
AD[23]  
V(I/O)  
AD[28]  
3.3V  
8
7
AD[29]  
GND  
6
5
BRSVP1A5  
BRSVP1A4  
INTA#  
BRSVP1B5  
GND  
RST#  
GND  
INTP  
5V  
4
V(I/O)  
INTC#  
TMS  
3
INTB#  
5V  
INTD#  
TDI  
2
TCK  
TDO  
+12V  
1
5V  
–12V  
TRST#  
5V  
PXI-1000 User Manual  
B-6  
© National Instruments Corporation  
 
   
Appendix B  
Pinouts  
Table B-6. P2 (J2) Connector Pinout for the Peripheral Slot  
Pin  
22  
21  
20  
19  
18  
17  
16  
15  
14  
13  
12  
11  
10  
9
Z
A
B
PXI_RSVB22  
GND  
C
PXI_RSVC22  
PXI_LBR1  
PXI_LBL0  
PXI_LBL3  
PXI_TRIG5  
PRST#  
D
PXI_RSVD22  
PXI_LBR2  
GND  
E
F
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
PXI_RSVA22  
PXI_LBR0  
PXI_LBR4  
PXI_LBL2  
PXI_TRIG3  
PXI_TRIG2  
PXI_TRIG1  
PXI_BRSVA15  
AD[35]  
PXI_RSVE22  
PXI_LBR3  
PXI_LBL1  
PXI_LBL5  
PXI_TRIG6  
PXI_CLK10  
PXI_TRIG7  
PXI_LBR6  
AD[32]  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
GND  
PXI_LBR5  
GND  
PXI_LBL4  
GND  
PXI_TRIG4  
GND  
PXI_STAR  
GND  
PXI_TRIG0  
GND  
DEG#  
FAL#  
PXI_LBL6  
GND  
AD[34]  
GND  
AD[33]  
AD[38]  
V(I/O)  
AD[37]  
AD[36]  
AD[42]  
AD[41]  
GND  
AD[40]  
GND  
AD[39]  
AD[45]  
V(I/O)  
AD[44]  
AD[43]  
AD[49]  
AD[48]  
GND  
AD[47]  
GND  
AD[46]  
AD[52]  
V(I/O)  
AD[51]  
AD[50]  
8
AD[56]  
AD[55]  
GND  
AD[54]  
GND  
AD[53]  
7
AD[59]  
V(I/O)  
AD[58]  
AD[57]  
6
AD[63]  
AD[62]  
GND  
AD[61]  
GND  
AD[60]  
5
C/BE[5]#  
V(I/O)  
V(I/O)  
C/BE[4]#  
GND  
PAR64  
4
PXI_BRSVB4  
GND  
C/BE[7]#  
PXI_LBR8  
SYSEN#  
PXI_LBL10  
C/BE[6]#  
PXI_LBR10  
PXI_LBL8  
PXI_LBL12  
3
PXI_LBR7  
PXI_LBR11  
PXI_LBL9  
PXI_LBR9  
PXI_LBL7  
PXI_LBL11  
2
PXI_LBR12  
GND  
1
© National Instruments Corporation  
B-7  
PXI-1000 User Manual  
 
   
C
Customer Communication  
For your convenience, this appendix contains forms to help you gather the information necessary  
to help us solve your technical problems and a form you can use to comment on the product  
documentation. When you contact us, we need the information on the Technical Support Form and  
the configuration form, if your manual contains one, about your system configuration to answer your  
questions as quickly as possible.  
National Instruments has technical assistance through electronic, fax, and telephone systems to quickly  
provide the information you need. Our electronic services include a bulletin board service, an FTP site,  
a fax-on-demand system, and e-mail support. If you have a hardware or software problem, first try the  
electronic support systems. If the information available on these systems does not answer your  
questions, we offer fax and telephone support through our technical support centers, which are staffed  
by applications engineers.  
Electronic Services  
Bulletin Board Support  
National Instruments has BBS and FTP sites dedicated for 24-hour support with a collection of files  
and documents to answer most common customer questions. From these sites, you can also download  
the latest instrument drivers, updates, and example programs. For recorded instructions on how to use  
the bulletin board and FTP services and for BBS automated information, call 512 795 6990. You can  
access these services at:  
United States: 512 794 5422  
Up to 14,400 baud, 8 data bits, 1 stop bit, no parity  
United Kingdom: 01635 551422  
Up to 9,600 baud, 8 data bits, 1 stop bit, no parity  
France: 01 48 65 15 59  
Up to 9,600 baud, 8 data bits, 1 stop bit, no parity  
FTP Support  
To access our FTP site, log on to our Internet host, ftp.natinst.com, as anonymous and use  
your Internet address, such as [email protected], as your password. The support files and  
documents are located in the /support directories.  
© National Instruments Corporation  
C-1  
PXI-1000 User Manual  
 
     
Fax-on-Demand Support  
Fax-on-Demand is a 24-hour information retrieval system containing a library of documents on a wide  
range of technical information. You can access Fax-on-Demand from a touch-tone telephone at  
512 418 1111.  
E-Mail Support (Currently USA Only)  
You can submit technical support questions to the applications engineering team through e-mail at the  
Internet address listed below. Remember to include your name, address, and phone number so we can  
contact you with solutions and suggestions.  
Telephone and Fax Support  
National Instruments has branch offices all over the world. Use the list below to find the technical  
support number for your country. If there is no National Instruments office in your country, contact  
the source from which you purchased your software to obtain support.  
Country  
Telephone  
Fax  
Australia  
Austria  
Belgium  
Brazil  
Canada (Ontario)  
Canada (Québec)  
Denmark  
Finland  
03 9879 5166  
0662 45 79 90 0  
02 757 00 20  
011 288 3336  
905 785 0085  
514 694 8521  
45 76 26 00  
09 725 725 11  
01 48 14 24 24  
089 741 31 30  
2645 3186  
03 6120092  
02 413091  
03 5472 2970  
02 596 7456  
5 520 2635  
03 9879 6277  
0662 45 79 90 19  
02 757 03 11  
011 288 8528  
905 785 0086  
514 694 4399  
45 76 26 02  
09 725 725 55  
01 48 14 24 14  
089 714 60 35  
2686 8505  
France  
Germany  
Hong Kong  
Israel  
Italy  
Japan  
03 6120095  
02 41309215  
03 5472 2977  
02 596 7455  
5 520 3282  
Korea  
Mexico  
Netherlands  
Norway  
Singapore  
Spain  
Sweden  
Switzerland  
Taiwan  
0348 433466  
32 84 84 00  
2265886  
91 640 0085  
08 730 49 70  
056 200 51 51  
02 377 1200  
01635 523545  
512 795 8248  
0348 430673  
32 84 86 00  
2265887  
91 640 0533  
08 730 43 70  
056 200 51 55  
02 737 4644  
01635 523154  
512 794 5678  
United Kingdom  
United States  
PXI-1000 User Manual  
C-2  
© National Instruments Corporation  
 
Technical Support Form  
Photocopy this form and update it each time you make changes to your software or hardware, and use  
the completed copy of this form as a reference for your current configuration. Completing this form  
accurately before contacting National Instruments for technical support helps our applications  
engineers answer your questions more efficiently.  
If you are using any National Instruments hardware or software products related to this problem,  
include the configuration forms from their user manuals. Include additional pages if necessary.  
Name __________________________________________________________________________  
Company _______________________________________________________________________  
Address ________________________________________________________________________  
_______________________________________________________________________________  
Fax ( ___ ) ________________Phone ( ___ ) __________________________________________  
Computer brand____________ Model ___________________Processor_____________________  
Operating system (include version number) ____________________________________________  
Clock speed ______MHz RAM _____MB  
Display adapter __________________________  
Mouse ___yes ___no Other adapters installed_______________________________________  
Hard disk capacity _____MB Brand_________________________________________________  
Instruments used _________________________________________________________________  
_______________________________________________________________________________  
National Instruments hardware product model _____________ Revision ____________________  
Configuration ___________________________________________________________________  
National Instruments software product ___________________ Version _____________________  
Configuration ___________________________________________________________________  
The problem is: __________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
List any error messages: ___________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
The following steps reproduce the problem: ___________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
 
PXI-1000 Hardware Configuration Form  
Record the settings and revisions of your hardware and software on the line to the right of each item.  
Complete a new copy of this form each time you revise your software or hardware configuration, and  
use this form as a reference for your current configuration. Completing this form accurately before  
contacting National Instruments for technical support helps our applications engineers answer your  
questions more efficiently.  
National Instruments Products  
Complete Part Number ____________________________________________________________  
Serial Number __________________________________________________________________  
Line Voltage ____________________________________________________________________  
Fan Speed ______________________________________________________________________  
Safety Ground Connected? _________________________________________________________  
Rack-Mount or Bench Top Configuration? ____________________________________________  
Using Remote Power Monitoring ____________________________________________________  
Filler Panels Installed in Empty Slots? ________________________________________________  
 
Other Products  
List and describe all devices installed in your mainframe.  
Slot  
Manufacturer, Description, and Function  
1
2
3
4
5
6
7
8
 
Documentation Comment Form  
National Instruments encourages you to comment on the documentation supplied with our products.  
This information helps us provide quality products to meet your needs.  
Title:  
PXI-1000 User Manual  
Edition Date: January 1998  
Part Number: 321710B-01  
Please comment on the completeness, clarity, and organization of the manual.  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
If you find errors in the manual, please record the page numbers and describe the errors.  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
_______________________________________________________________________________  
Thank you for your help.  
Name _________________________________________________________________________  
Title __________________________________________________________________________  
Company _______________________________________________________________________  
Address ________________________________________________________________________  
_______________________________________________________________________________  
E-Mail Address __________________________________________________________________  
Phone ( ___ ) __________________________ Fax ( ___ ) _______________________________  
Mail to: Technical Publications  
National Instruments Corporation  
Fax to: Technical Publications  
National Instruments Corporation  
512 794 5678  
6504 Bridge Point Parkway  
Austin, Texas 78730-5039  
 
Glossary  
Prefix  
n-  
Meanings  
nano-  
Value  
109  
10– 6  
103  
102  
103  
µ-  
micro-  
milli-  
m-  
c-  
centi-  
k-  
kilo-  
M-  
mega-  
106  
Symbols  
°
Degrees.  
%
Equal or greater than.  
Equal or less than.  
Percent.  
A
A
Amperes.  
AC  
Alternating current.  
ANSI  
AWG  
American National Standards Institute.  
American Wire Gauge.  
B
backplane  
An assembly, typically a printed circuit board, with connectors and signal  
paths that bus the connector pins.  
© National Instruments Corporation  
G-1  
PXI-1000 User Manual  
 
   
Glossary  
C
C
Celsius.  
cfm  
CFR  
CSA  
Cubic feet per minute.  
Cooperative Fuel Research.  
Canadian Standards Association.  
D
daisy-chain  
A method of propagating signals along a bus, in which the devices are  
prioritized on the basis of their position on the bus.  
DC  
Direct current.  
E
ECL  
EIA  
EMC  
Emitter-coupled logic.  
Electronic Industries Association.  
Electromagnetic Compatibility.  
F
FCC  
Federal Communications Commission.  
G
g
1) grams 2) A measure of acceleration equal to 9.8 m/s2.  
General Purpose Interface Bus (IEEE 488).  
GPIB  
gRMS  
A measure of random vibration. The root mean square of acceleration  
levels in a random vibration test profile.  
PXI-1000 User Manual  
G-2  
© National Instruments Corporation  
 
Glossary  
H
Hz  
Hertz; cycles per second.  
I
IEC  
International Electrotechnical Commission; an organization that sets  
international electrical and electronics standards.  
IEEE  
IMP  
Institute of Electrical and Electronics Engineers.  
Mainframe peak current.  
Inches.  
in.  
L
lb  
Pounds.  
M
m
Meters.  
MTBF  
MTTR  
Mean time between failure.  
Mean time to repair.  
N
NEMA  
National Electrical Manufacturers Association.  
PCI eXtensions for Instrumentation.  
Relative humidity.  
P
PXI  
R
RH  
© National Instruments Corporation  
G-3  
PXI-1000 User Manual  
 
Glossary  
RMS  
Root mean square. A method used to measure electrical output in volts and  
watts.  
S
s
Seconds.  
ST  
Star Trigger.  
Star Trigger slot  
This slot is located at slot 2 and has a dedicated trigger line between each  
peripheral slot. Use this slot for a module with ST functionality that can  
provide individual triggers to all other peripherals.  
System controller  
A module configured for installation in Slot 0 of a VXIbus mainframe. This  
device is unique in the VXIbus system in that it performs the VMEbus  
system controller functions, including clock sourcing and arbitration for  
data transfers across the backplane. Installing such a device into any other  
slot can damage the device, the VXIbus backplane, or both.  
U
UL  
Underwriter’s Laboratories.  
V
V
Volts.  
VPP  
VAC  
Peak to peak voltage.  
Volts alternating current.  
W
W
Watts.  
PXI-1000 User Manual  
G-4  
© National Instruments Corporation  
 
Index  
A
D
AC input (table), A-1  
AC mains power  
DB-9 connector  
pinout (table), 2-5  
connecting to, 2-3 to 2-4  
resetting circuit breaker, 3-3  
air cooling. See cooling.  
power supply voltages (table), 2-4  
DC output (table), A-2  
documentation  
conventions used in manual, xi  
organization of manual, xi  
related documentation, xii  
B
bulletin board support, C-1  
E
C
electrical specifications, A-1 to A-2  
AC input (table), A-1  
cables, power (table), 1-2  
chassis initialization file, 2-7  
circuit breaker, resetting, 3-3  
cleaning  
DC output (table), A-2  
electronic support services, C-1 to C-2  
e-mail support, C-2  
exterior cleaning, 3-2  
fan filters, 3-2  
environmental specifications (table), A-4  
equipment, optional, 1-2  
interior cleaning, 3-2  
precautions before cleaning, 3-1  
CompactPCI  
F
fan  
installing modules (figure), 2-6  
interoperability with PXI-1000 backplane,  
1-3  
cleaning fan filters, 3-2  
setting speed, 2-2  
configuration. See installation and configuration.  
connector pinouts. See pinouts.  
cooling  
speed selector switch (figure), 1-5  
fax and telephone support, C-2  
Fax-on-Demand support, C-2  
filler panel installation, 2-7  
FTP support, C-1  
air cooling of PXI-1000, 2-1 to 2-2  
air intake (figure), 2-2  
filler panel installation, 2-7  
setting fan speed, 2-2  
specifications (table), A-3  
customer communication, xii, C-1 to C-2  
G
ground, connecting, 2-3  
© National Instruments Corporation  
I-1  
PXI-1000 User Manual  
 
   
Index  
precautions before cleaning, 3-1  
preparation, 3-1  
resetting AC mains circuit breaker, 3-3  
service interval, 3-1  
static discharge damage (caution), 3-1  
troubleshooting, 3-4  
I
Inhibit signal, turning off power supply  
outputs, 2-5  
installation and configuration, 2-1 to 2-7  
chassis initialization file, 2-7  
connecting safety ground, 2-3  
connecting to AC mains power, 2-3 to 2-4  
filler panel installation, 2-7  
module installation, 2-6 to 2-7  
CompactPCI or PXI module  
(figure), 2-6  
mechanical specifications, A-6 to A-7  
O
optional equipment, 1-2  
ejector/injector handle position  
(figure), 2-7  
rack mounting, 2-2  
P
P1 (J1) connector pinouts  
peripheral slot (table), B-6  
Star Trigger slot (table), B-4  
System Controller slot (table), B-2  
P2 (J2) connector pinouts  
peripheral slot (table), B-7  
Star Trigger slot (table), B-5  
System Controller slot (table), B-3  
peripheral slots  
remote power monitoring and inhibit  
interface, 2-5  
setting fan speed, 2-2  
site considerations, 2-1 to 2-2  
testing power up, 2-3 to 2-4  
unpacking the PXI-1000, 1-1  
J
overview, 1-5  
J1 connector pinouts. See P1 (J1) connector  
P1 (J1) connector pinouts (table), B-6  
P2 (J2) connector pinouts (table), B-7  
pinouts  
pinouts.  
J2 connector pinouts. See P2 (J2) connector  
pinouts.  
DB-9 connector (table), 2-5  
P1 (J1) connector  
peripheral slot (table), B-6  
Star Trigger slot (table), B-4  
System Controller slot (table), B-2  
P2 (J2) connector  
L
local bus  
routing (figure), 1-6  
signals, 1-6  
peripheral slot (table), B-7  
Star Trigger slot (table), B-5  
System Controller slot (table), B-3  
power cables (table), 1-1 to 1-2  
power monitoring connector. See DB-9  
connector.  
M
maintenance of PXI-1000, 3-1 to 3-4  
cleaning  
exterior cleaning, 3-2  
fan filters, 3-2  
power problems, troubleshooting, 3-4  
interior cleaning, 3-2  
PXI-1000 User Manual  
I-2  
© National Instruments Corporation  
 
Index  
power supply  
remote power monitoring and inhibit  
S
safety ground, connecting, 2-3  
safety specifications (table), A-3  
service interval, 3-1  
specifications, A-1 to A-7  
backplane, A-5  
interface, 2-5  
universal, 2-3  
voltages at power monitoring connector  
(DB-9) (table), 2-4  
power up, testing, 2-3 to 2-4  
PXI-1000  
cooling, A-3  
electrical, A-1 to A-2  
environmental, A-4  
mechanical, A-7  
cooling air intake (figure), 2-2  
dimensions (figure), A-7  
features, 1-2 to 1-3  
front view (figure), 1-4  
installation (See installation and  
configuration)  
safety, A-3  
Star Trigger (ST) slot  
description, 1-5  
P1 (J1) connector pinouts (table), B-4  
P2 (J2) connector pinouts (table), B-5  
triggers for peripheral slots (figure), 1-6  
static discharge damage (caution), 3-1  
System Controller slot  
maintenance (See maintenance of  
PXI-1000)  
optional equipment, 1-2  
power cables required (table), 1-1 to 1-2  
rear view (figure), 1-5  
requirements for getting started,  
1-1 to 1-2  
description, 1-5  
P1 (J1) connector pinouts (table), B-2  
P2 (J2) connector pinouts (table), B-3  
system reference clock, 1-7  
specifications (See specifications)  
unpacking, 1-1  
PXI-1000 backplane, 1-3 to 1-7  
interoperability with CompactPCI, 1-3  
local bus, 1-6  
T
technical support, C-1 to C-2  
telephone and fax support, C-2  
trigger bus, 1-6 to 1-7  
peripheral slots, 1-5  
specifications (table), A-5  
Start Trigger slot, 1-5  
troubleshooting the PXI-1000, 3-4  
System Controller slot, 1-5  
system reference clock, 1-7  
PXI_CLK10, 1-7  
U
PXI_CLK10_IN pin, 1-7  
unpacking the PXI-1000, 1-1  
R
V
rack mounting, 2-2  
voltages at power monitoring connector  
(DB-9) (table), 2-4  
remote power monitoring and inhibit  
interface, 2-5  
© National Instruments Corporation  
I-3  
PXI-1000 User Manual  
 

Miele Freezer F1801SF User Manual
NETGEAR Network Router DGN1000 100NAS User Manual
Nokia Portable Multimedia Player CK 20W User Manual
NXP Semiconductors Switch PIP3107 D User Manual
Oster Toaster TSSTTVVG01 User Manual
Panasonic All in One Printer KX MB2060 User Manual
Panasonic Car Stereo System CQ 5251JU User Manual
Panasonic GPS Receiver CN GP50N User Manual
Parasound Speaker CS T 255A User Manual
Patton electronic IP Phone SmartNode 4110 Series User Manual