National Instruments Network Card PXI PXITM 1000 User Manual |
Important Information
Warranty
The PXI-1000 is warranted against defects in materials and workmanship for a period of one year from the date of
shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace
equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming
instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced
by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not
execute programming instructions if National Instruments receives notice of such defects during the warranty period.
National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside
of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs
of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this manual is accurate. The document has been carefully reviewed
for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to
make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should
consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages
arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS
ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED
BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE
CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS,
OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of the liability of
National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence.
Any action against National Instruments must be brought within one year after the cause of action accrues. National
Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty
provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow
the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product;
owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties,
or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical,
including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without
the prior written consent of National Instruments Corporation.
Trademarks
PXI™ is a trademark of National Instruments Corporation.
Product and company names listed are trademarks or trade names of their respective companies.
WARNING REGARDING MEDICAL AND CLINICAL USE OF NATIONAL INSTRUMENTS PRODUCTS
National Instruments products are not designed with components and testing intended to ensure a level of reliability
suitable for use in treatment and diagnosis of humans. Applications of National Instruments products involving medical
or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the part of the
user or application designer. Any use or application of National Instruments products for or involving medical or clinical
treatment must be performed by properly trained and qualified medical personnel, and all traditional medical safeguards,
equipment, and procedures that are appropriate in the particular situation to prevent serious injury or death should always
continue to be used when National Instruments products are being used. National Instruments products are NOT intended
to be a substitute for any form of established process, procedure, or equipment used to monitor or safeguard human health
and safety in medical or clinical treatment.
Compliance
FCC/DOC Radio Frequency Interference
Class A Compliance
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance
with the instructions in this manual, may cause interference to radio and television reception. Classification
requirements are the same for the Federal Communications Commission (FCC) and the Canadian
Department of Communications (DOC). This equipment has been tested and found to comply with the
following two regulatory agencies:
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
Notices to User: Changes or modifications not expressly approved by National Instruments could void
the user’s authority to operate the equipment under the FCC Rules.
This device complies with the FCC rules only if used with shielded interface cables
of suitable quality and construction. National Instruments used such cables to test
this device and provides them for sale to the user. The use of inferior or nonshielded
interface cables could void the user’s authority to operate the equipment under the
FCC rules.
If necessary, consult National Instruments or an experienced radio/television technician for additional
suggestions. The following booklet prepared by the FCC may also be helpful: Interference to Home
Electronic Entertainment Equipment Handbook. This booklet is available from the U.S. Government
Printing Office, Washington, DC 20402.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment
Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur
du Canada.
For Your Safety
Caution
Before undertaking any troubleshooting, maintenance, or exploratory procedure,
read carefully the WARNING and CAUTION notices.
!
capable of inflicting personal injury.
•
•
Mainframe Grounding—The PXI-1000 mainframe requires a
connection from the premise wire safety ground to the PXI-1000
chassis ground. The earth safety ground must be connected during use
of this equipment to minimize shock hazards. Refer to the Connecting
Safety Ground section of Chapter 2, Installation and Configuration,
for instructions on connecting safety ground.
Live Circuits—Operating personnel and service personnel must not
remove protective covers when operating or servicing the PXI-1000.
Adjustments and service to internal components must be undertaken
by qualified service technicians. During service of this product, the
mains connector to the premise wiring must be disconnected.
Dangerous voltages may be present under certain conditions; use
extreme caution.
•
•
Explosive Atmosphere—Do not operate the mainframe in conditions
where flammable gases are present. Under such conditions this
equipment is unsafe and may ignite the gases or gas fumes.
Part Replacement—Only service this equipment with parts that are
exact replacements, both electrically and mechanically. Contact
National Instruments for replacement part information. Installation of
parts with those that are not direct replacements may cause harm to
personnel operating the mainframe. Furthermore, damage or fire may
occur if replacement parts are unsuitable.
•
Modification—Do not modify any part of the mainframe from its
original condition. Unsuitable modifications may result in safety
hazards.
About This Manual
Organization of This Manual.........................................................................................xi
Conventions Used in This Manual.................................................................................xi
Chapter 1
Getting Started
Unpacking......................................................................................................................1-1
Optional Equipment.......................................................................................................1-2
PXI-1000 Backplane Overview.....................................................................................1-3
System Controller Slot ....................................................................................1-5
Star Trigger Slot ..............................................................................................1-5
Local Bus.........................................................................................................1-6
Chapter 2
Installation and Configuration
Connecting Safety Ground.............................................................................................2-3
Connecting to AC Mains Power and Testing Power up ................................................2-3
Installing PXI Modules..................................................................................................2-6
Chapter 3
Maintenance
Service Interval..............................................................................................................3-1
Preparation.....................................................................................................................3-1
© National Instruments Corporation
vii
PXI-1000 User Manual
Contents
Cleaning......................................................................................................................... 3-1
Interior Cleaning ............................................................................................. 3-2
Cleaning the Fan Filters................................................................................................. 3-2
Resetting the AC Mains Circuit Breaker....................................................................... 3-3
Troubleshooting the PXI-1000...................................................................................... 3-4
Appendix A
Specifications
Appendix B
Pinouts
Appendix C
Customer Communication
Glossary
Index
Figures
Figure 1-1.
Figure 1-2.
Figure 1-3.
Rear View of the PXI-1000 Mainframe................................................ 1-5
Figure 2-1.
Figure 2-2.
Figure 2-3.
PXI-1000 Mainframe Airflow Side View............................................. 2-2
Injector/Ejector Handle Position during Module Insertion................... 2-7
Figure A-1. PXI-1000 Dimensions........................................................................... A-7
Tables
Table 1-1.
Power Cables ........................................................................................ 1-1
Table 2-1.
Table 2-2.
Power Supply Voltages at Power Monitoring Connector (DB-9) ........ 2-4
DB-9 Connector Pinout......................................................................... 2-5
Table 3-1.
Troubleshooting .................................................................................... 3-4
PXI-1000 User Manual
viii
© National Instruments Corporation
Table A-1.
Table A-2.
Table A-3.
Table A-4.
Table A-5.
Table A-6.
Table A-7.
DC Output Specifications......................................................................A-2
Cooling Specifications ..........................................................................A-3
Table B-1.
Table B-2.
Table B-3.
Table B-4.
Table B-5.
Table B-6.
P1 (J1) Connector Pinout for the System Controller Slot .....................B-2
P2 (J2) Connector Pinout for the System Controller Slot .....................B-3
P1 (J1) Connector Pinout for the Star Trigger Slot...............................B-4
P2 (J2) Connector Pinout for the Star Trigger Slot...............................B-5
P1 (J1) Connector Pinout for the Peripheral Slot..................................B-6
P2 (J2) Connector Pinout for the Peripheral Slot..................................B-7
© National Instruments Corporation
ix
PXI-1000 User Manual
About This Manual
The PXI-1000 User Manual describes the features of the PXI-1000
mainframe and contains information about configuring the mainframe,
Organization of This Manual
This manual is organized as follows:
•
lists the contents of your kit, and lists optional equipment you can
•
•
•
•
•
Chapter 2, Installation and Configuration, describes how to prepare
Chapter 3, Maintenance, describes basic maintenance procedures you
Appendix A, Specifications, contains complete specifications for the
PXI-1000 mainframe.
the PXI-1000 backplane.
request help from National Instruments or to comment on our products
and manuals.
•
•
The Glossary lists abbreviations, acronyms, metric prefixes,
mnemonics, symbols, and terms.
The Index contains an alphabetical list of key terms and topics used in
this manual, including the page where you can find each one.
Conventions Used in This Manual
The following conventions are used in this manual:
This icon to the left of bold italicized text denotes a note, which alerts you
to important information.
This icon to the left of bold italicized text denotes a caution, which advises
you of precautions to take to avoid injury, data loss, or a system crash.
!
© National Instruments Corporation
xi
PXI-1000 User Manual
About This Manual
This icon to the left of bold italicized text denotes a warning, which advises
you of precautions to take to avoid being electrically shocked.
bold italic
Bold italic text denotes a note, caution, or warning.
italic
Italic text denotes emphasis, a cross reference, or an introduction to a key
concept. This font also denotes text from which you supply the appropriate
word or value, as in Windows 3.x.
monospace
Text in this font denotes text or characters that are to be literally enter from
the keyboard, sections of code, programming examples, and syntax
examples. This font is also used for the proper names of disk drives, paths,
directories, device names, functions, variables, filenames, and extensions.
Related Documentation
The following documents contain information that you may find helpful as
you read this manual:
•
•
•
Compact PCI Specification PICMG 2.0 R2.1
PXI Specification Revision 1.0
IEEE 1101.1-1991, IEEE Standard for Mechanical Core
Specifications for Microcomputers Using IEC 603-2 Connectors
•
IEEE 1101.10 and P1101.11, IEEE Standard for Additional
Mechanical Specifications for Microcomputers Using IEEE 1101.1
Equipment Practice
Customer Communication
National Instruments wants to receive your comments on our products
and manuals. We are interested in the applications you develop with our
products, and we want to help if you have problems with them. To make it
easy for you to contact us, this manual contains comment and configuration
forms for you to complete. These forms are in Appendix C, Customer
Communication, at the end of this manual.
PXI-1000 User Manual
xii
© National Instruments Corporation
1
Getting Started
This chapter describes the key features of the PXI-1000, lists the contents
of your kit, and lists optional equipment you can order from National
Instruments.
Unpacking
Carefully inspect the shipping container and the mainframe for damage.
Check for visible damage to the metal work. Check to make sure all
handles, hardware, and switches are undamaged. Inspect the inner chassis
for any possible damage, debris, or detached components. If damage
appears to have been caused in shipment, file a claim with the carrier.
Retain the packing material for possible inspection and/or reshipment.
What You Need to Get Started
The PXI-1000 kit contains the following items:
❑ PXI-1000 mainframe
❑ Filler panels
❑ Power cable (see Table 1-1)
❑ PXI-1000 User Manual
❑ Floppy disk with Chassis Initialization file, chassis.ini
Table 1-1. Power Cables
Power Cable
Standard 120 V (USA)
Switzerland 220 V
Australia 240 V
Reference Standards
ANSI C73.11/NEMA 5-15-P/IEC83
SEV
AS C112
© National Instruments Corporation
1-1
PXI-1000 User Manual
Chapter 1
Getting Started
Table 1-1. Power Cables (Continued)
Power Cable Reference Standards
Universal Euro 240 V
North America 240 V
United Kingdom 240 V
CEE (7), II, IV, VII IEC83
ANSI C73.20/NEMA 5-15-P/IEC83
BS 1363/IEC83
If you are missing any of the above items or if you have the incorrect power
cord, contact National Instruments.
Optional Equipment
An optional rack-mount kit is available from National Instruments. You
can use this kit to install the PXI-1000 mainframe into a standard 19 in.
(482 mm) instrument cabinet. Contact National Instruments to order the
PXI-1000 rack-mount kit.
Key Features
The PXI-1000 combines a high-performance 8-slot PXI backplane with a
high-output power supply and a structural design that has been optimized
for maximum usability in a wide range of applications. The mainframe’s
modular design ensures the highest level of maintainability resulting in a
very low mean time to repair (MTTR). The PXI-1000 fully complies with
the PXI Specification, Revision 1.0, offering advanced timing and
synchronization features.
The key features of the PXI-1000 include:
•
•
•
•
•
PXI and CompactPCI (PICMG 2.0 R 2.1) module compatibility
Compact 3U-sized, 8-slot chassis
300 W of usable power
Universal AC input: auto-voltage and auto-frequency ranging
Over-current protection via push-reset circuit breaker (no loose fuses
to replace)
•
•
•
Removable modular power supply with MTBF of 90,000 hours
Remote power status and inhibit via a rear panel connector
On/Off (Standby) switch located on the front panel for easy access
PXI-1000 User Manual
1-2
© National Instruments Corporation
Chapter 1
Getting Started
•
•
•
Selectable fan speed for maximum cooling or quiet operation
Carrying handle for portability
Tilt feet for bench-top applications
PXI-1000 Backplane Overview
Interoperability with CompactPCI
The PXI-1000 backplane is interoperable with PXI-compatible products
and standard CompactPCI products. This is an important feature, as many
PXI-compatible systems may not require components that do not
implement PXI-specific features. For example, you may want to use a
standard CompactPCI network interface card in a PXI chassis.
The signals on the P1 connector of the backplane meet the requirements of
the CompactPCI specification for both the peripheral and system modules.
specification. Therefore, all modules that meet the requirements of the
CompactPCI 64-bit specification will function in the PXI-1000.
Figures 1-1 and 1-2 show some of the key features and components of the
PXI-1000 mainframe. Figure 1-1 shows the front view of the PXI-1000 and
Figure 1-2 shows the rear view.
© National Instruments Corporation
1-3
PXI-1000 User Manual
Chapter 1
Getting Started
1
ON
STANDBY
22
P2
1
25
2
P1
1
ZABCDEF
2
3
4
5
6
7
8
1
5
3
4
8
7
6
4
3
1
2
Remote On/Off (Standby) Switch 3 Optional Mounting Brackets
6
7
8
Star Trigger Slot
Backplane Connectors
(Located in Slots 1-8)
4
5
Removable Feet
Peripheral Slots
System Controller Slot
Controller Expansion Slots
Figure 1-1. Front View of the PXI-1000 Mainframe
PXI-1000 User Manual
1-4
© National Instruments Corporation
Chapter 1
Getting Started
1
2
3
90-264 VAC
47-63 HZ
8A MAX
CB1
10A
DISCONNECT
POWEORD
BEFORE REMOVING
POWER MODULE
HI
LO
FAN SPEED
DB-9
7
6
5
4
1
2
Module Cooling Air Outlets
Universal AC Input
3
4
Main Power Switch
Circuit Breaker
5
Chassis Ground Screw
7
Fan Speed
Selector
6
Remote Power Status and
Inhibit Connector
Figure 1-2. Rear View of the PXI-1000 Mainframe
System Controller Slot
The System Controller slot is located in Slot 1 of the chassis as defined by
the PXI specification. It has three controller expansion slots, which are
used for system controller modules that are wider than one slot. As defined
in the PXI specification, these slots allow the controller to expand to the left
to prevent the controller from using up peripheral slots.
Star Trigger Slot
The Star Trigger (ST) slot is located at Slot 2. This slot has a dedicated
trigger line between each peripheral slot (see Figure 1-3). This slot is
intended for modules with ST functionality that can provide individual
triggers to all other peripherals. However, if you do not require advanced
trigger functionality, you can install any standard peripheral module into
this slot.
Peripheral Slots
There are seven peripheral slots including the Star Trigger controller slot.
© National Instruments Corporation
1-5
PXI-1000 User Manual
Chapter 1
Getting Started
Local Bus
The PXI backplane’s local bus is a daisy-chained bus that connects each
peripheral slot with its adjacent peripheral slots to the left and right, as
shown in Figure 1-3.
For example, a given peripheral slot’s right local bus connects to the
adjacent slot’s left local bus and so on. Each local bus is 13 lines wide and
can pass analog signals between cards or provide a high-speed side-band
communication path that does not affect the PXI bandwidth.
Local Bus signals may range from high-speed TTL signals to analog
signals as high as 42 V. Initialization software keys adjacent boards to
prohibit the use of incompatible boards. This software uses the
configuration information specific to each peripheral board to evaluate
compatibility. This method is a flexible way to define local bus
functionality that is not limited by hardware keying.
Star Triggers
Local
Bus
Local
Bus
Local
Bus
PCI Arbitration and Clock Signals
Figure 1-3. PXI Local Bus and Star Trigger Routing
Trigger Bus
The eight PXI trigger lines are bused to each slot. You can use the trigger
lines in a variety of ways. For example, you can use triggers to synchronize
the operation of several different PXI peripheral modules. In other
applications, one module can control carefully timed sequences of
PXI-1000 User Manual
1-6
© National Instruments Corporation
Chapter 1
Getting Started
operations performed on other modules in the system. Modules can pass
triggers to one another, allowing precisely timed responses to
asynchronous external events the system is monitoring or controlling.
System Reference Clock
The PXI-1000 supplies the PXI 10 MHz system clock signal (PXI_CLK10)
independently to each peripheral slot. An independent buffer (having a
between slots) drives the clock signal to each peripheral slot. You can use
this common reference clock signal to synchronize multiple modules in a
measurement or control system. You can drive PXI_CLK10 from an
external source through the PXI_CLK10_IN pin on the P2 connector of the
Star Trigger Slot. (See Table B-1, P1 (J1) Connector Pinout for the System
Controller Slot, in Appendix B, Pinouts.) Sourcing an external clock on
this pin automatically disables the backplane’s 10 MHz source.
© National Instruments Corporation
1-7
PXI-1000 User Manual
2
Installation and Configuration
This chapter describes how to prepare and operate your PXI-1000
mainframe.
Before connecting the mainframe to a power source, read this chapter and
the For Your Safety section located at the beginning of this manual.
Site Considerations
The PXI-1000 is designed to operate on a bench or in an instrument rack.
installation instructions.
Apertures in the rear and along both sides of the mainframe facilitate power
supply and module cooling. Air enters through filters and fan inlets located
in the lower rear of the mainframe and exits through the upper sections on
both sides and through the rear, as shown in Figure 2-1. Place your
PXI-1000 on a bench top or in an instrument rack so that the fans (air inlets)
and the air outlet apertures along both sides of the mainframe have adequate
ventilation. Keep other equipment a minimum of 3.0 in. (76.2 mm) away
from the air inlets and outlets.
© National Instruments Corporation
2-1
PXI-1000 User Manual
Chapter 2
Installation and Configuration
1
2
1
Module Cooling Air Outlets
2
Module Cooling Air Intake
Figure 2-1. PXI-1000 Mainframe Airflow Side View
Install your mainframe so that you can easily access the rear panel. This
simplifies the replacement of the air filters or power supply/fan assembly,
if necessary.
Rack Mounting
Rack-mount applications require the optional rack-mount kit available
from National Instruments. Refer to the instructions supplied with the
rack-mount kit to install your PXI-1000 in an instrument rack.
Note
You may wish to remove the feet from your PXI-1000 when rack mounting. To do
so, remove the two screws holding the feet in place.
Setting Fan Speed
The fan speed selector switch is located on the rear panel of the PXI-1000.
Refer to Figure 1-2, Rear View of the PXI-1000 Mainframe, for a diagram
of the fan speed selector. Select HI for maximum cooling effectiveness
(recommended) or LO for more quiet operation.
PXI-1000 User Manual
2-2
© National Instruments Corporation
Chapter 2
Installation and Configuration
Connecting Safety Ground
Warning The PXI-1000 chassis is designed with a three-position NEMA 15-5 style plug that
connected whenever you power up the chassis.
If your power outlet does not have an appropriate ground connection, you
must connect the premise wire safety ground to the chassis grounding
screw located on the rear panel. Refer to Figure 1-2, Rear View of the
PXI-1000 Mainframe, for a diagram of the chassis grounding screw. To
connect the safety ground, complete the following steps:
1. Connect a 16 AWG (1.3 mm) wire to the chassis grounding screw
using a toothed grounding lug. The wire must have green insulation
with a yellow stripe or must be non-insulated (bare).
2. Attach the opposite end of the wire to permanent earth ground using
toothed washers or a toothed lug.
Connecting to AC Mains Power and Testing Power up
Caution
Make sure the main power switch located on the rear panel of the PXI-1000 is in
a diagram of the main power switch.
!
The power supply is universal, which means the mainframe can connect to
all standard worldwide input voltages. Refer to Chapter 1, Getting Started,
for power cord specifications. Attach input power through the rear AC inlet
using the appropriate line cord supplied. Refer to Figure 1-2, Rear View of
the PXI-1000 Mainframe, for a diagram of the IEC 320 inlet.
Push the main power switch to the On (1) position. Then push the remote
power switch to the On position (if not already on). Observe that all fans
become operational. Note that the main power switch (rear panel) enables
the power module to receive AC power. The remote front panel standby
switch allows the power module to provide power to the PXI backplane
(if the rear main power switch is on). For this reason, you can leave the
main power switch on and use the remote On/Off (Standby) switch on the
front to power up and power down the backplane.
© National Instruments Corporation
2-3
PXI-1000 User Manual
Chapter 2
Installation and Configuration
Caution
When connecting digital voltmeter probes to the rear D-sub connector, be careful
not to short the probe leads together. Doing so could damage the power supply.
!
PXI-1000 are within the allowable limits. Referring to Table 2-1, connect
one lead of the voltmeter to a supply pin on the remote power monitoring
connector (9-pin D-sub) located on the rear panel. Refer to Table 2-2 for a
pinout diagram of the remote power monitoring connector. Connect the
reference lead of the voltmeter to one of the ground pins. Compare each
voltage reading to the values listed in Table 2-1.
Note
Use the rear panel D-sub connector to check voltages only. Do not use the
connector to supply power to external devices.
Table 2-1. Power Supply Voltages at Power Monitoring Connector (DB-9)
Pin
2
Supply
+5 V
Acceptable Voltage Range
4.75 to 5.25 V
3.135 to 3.465 V
11.4 to 12.6 V
–12.6 to –11.4 V
N/A
4
+3.3 V
6
+12 V
8
–12 V
1, 9
Logic Ground
with the CompactPCI voltage limit specifications. Note that the rear panel
D-sub connector is to be used to check voltages only. Do not use these
voltages to supply power to external devices.
Note
If the fans or power unit fail to function properly, refer to the Troubleshooting the
PXI-1000 section of Chapter 3, Maintenance.
PXI-1000 User Manual
2-4
© National Instruments Corporation
Chapter 2
Installation and Configuration
Remote Power Monitoring and Inhibiting Interface
The PXI-1000 mainframe supports remote power monitoring and
inhibiting via a 9-pin D-sub connector located on the rear panel. Table 2-2
shows the pinout of the DB-9 connector.
Table 2-2. DB-9 Connector Pinout
DB-9 Pin
Signal
1
2
3
4
5
6
7
8
9
Logic Ground
+5 V
Reserved
+3.3 V
Inhibit*
+12 V
Reserved
–12 V
Logic Ground
5
4
3
2
1
9
8
7
6
You can use the Inhibit signal (active low) to turn off the power supply
outputs. To use this feature, connect the Inhibit pin (pin 5) to a Logic
Ground pin (pin 1 or 9). Make sure both the rear main power switch and
the front (standby) switch are in the ON position. As long as Inhibit is
pulled down, the power supply inhibits its DC outputs. DC output resumes
when Inhibit is no longer connected to ground. Note that the power
(standby) switch, located on the front of the chassis, uses this inhibiting
feature. For remote reset, connect a momentary switch between pin 5 and
pin 1 (or pin 9).
© National Instruments Corporation
2-5
PXI-1000 User Manual
Chapter 2
Installation and Configuration
Installing PXI Modules
Caution
Turn off the mainframe power before installing CompactPCI or PXI modules.
!
Install a module into a mainframe slot by first placing the module’s card
edges into the front module guides (top and bottom), as shown in
Figure 2-2. Slide the module to the rear of the mainframe (making sure that
the injector/ejector handle is pushed down as shown in Figure 2-3). When
you begin to feel resistance, push up on the injector/ejector handle to inject
the card into the frame. Secure the module’s front panel to the mainframe
using the module’s front-panel mounting screws.
PXI-1000 Chassis
PXI-8156
System Controller
ON
STANDBY
1
2
3
4
5
6
7
8
PXI Board
Injector/Ejector Rail
Injector/Ejector Handle
Figure 2-2. Installing PXI or CompactPCI Modules
PXI-1000 User Manual
2-6
© National Instruments Corporation
Chapter 2
Installation and Configuration
Figure 2-3. Injector/Ejector Handle Position during Module Insertion
Installing Filler Panels
To optimize module cooling performance, install filler panels into unused
or empty slots. Secure with the captive mounting screws.
Using the Chassis Initialization File
To assist system integrators, the PXI specification requires manufacturers
of PXI chassis and system modules to document the capabilities of their
products. The minimum documentation requirements are contained in
.ini files, which consist of ASCII text. The system integrator can read the
.ini file, and configuration utilities and device drivers can also use this
file. The PXI-1000 chassis initialization file, chassis.ini, is included on
the diskette for your PXI-1000.
© National Instruments Corporation
2-7
PXI-1000 User Manual
3
Maintenance
This chapter describes basic maintenance procedures you can perform on
the PXI-1000 mainframe.
Service Interval
Clean the mainframe fan filters at a maximum interval of six months.
Depending upon the amount of use and ambient dust levels in the operating
environment, the filters may require more frequent cleaning.
Clean dust from the mainframe exterior (and interior) as needed, based on
Preparation
The information in this section is designed for use by qualified service
personnel. Read the For Your Safety section at the beginning of this manual
before attempting any procedures in this chapter.
Caution
Many components within the mainframe are susceptible to static discharge
damage. Service the mainframe only in a static-free environment. Observe
standard handling precautions for static-sensitive devices while servicing the
mainframe. Always wear a grounded wrist strap, or equivalent, while servicing
the mainframe.
!
Cleaning
Cleaning procedures consist of exterior and interior cleaning of the
mainframe and cleaning the fan filters. Refer to your module user
documentation for information on cleaning the individual CompactPCI
or PXI modules.
Caution
Always power-off the mainframe (using the main power switch on the rear of the
mainframe) and disconnect the power cord before cleaning or servicing the
mainframe.
!
© National Instruments Corporation
3-1
PXI-1000 User Manual
Chapter 3
Maintenance
Interior Cleaning
Exterior Cleaning
Use a dry, low-velocity stream of air to clean the interior of the mainframe.
Use a soft-bristle brush for cleaning around components. If you must use a
liquid for minor interior cleaning, use a 75% isopropyl alcohol solution and
rinse with deionized water.
Clean the exterior surfaces of the mainframe with a dry lint-free cloth or a
soft-bristle brush. If any dirt remains, wipe with a cloth moistened in a mild
soap solution. Remove any soap residue by wiping with a cloth moistened
with clear water. Do not use abrasive compounds on any part of the
mainframe.
Cautions Avoid getting moisture inside the mainframe during exterior cleaning. Use just
!
enough moisture to dampen the cloth.
Do not wash the front- or rear-panel connectors or switches. Cover these
components while cleaning the mainframe.
Do not use chemical cleaning agents; they may damage the mainframe. Avoid
chemicals that contain benzene, toluene, xylene, acetone, or similar solvents.
Cleaning the Fan Filters
You can easily remove the mainframe cooling filters from the rear of the
mainframe by removing the plastic housing attached to each fan.
Clean the fan filters by washing them in a mild soap solution and then
vacuuming or blowing air through them. Rinse the filters with water and
allow them to dry before reinstalling them on the mainframe.
PXI-1000 User Manual
3-2
© National Instruments Corporation
Chapter 3
Maintenance
Resetting the AC Mains Circuit Breaker
If your PXI-1000 encounters an over-current condition, the circuit breaker
located on the rear panel will trip to prevent damage to the mainframe.
Complete the following steps to reset the circuit breaker:
1. Turn the front panel (remote) power switch to the Standby position.
2. Turn off the rear main power switch.
3. Depress the circuit breaker to reset it.
4. Turn on the rear main power switch.
5. Turn the front panel (remote) power switch to the On position.
If the circuit breaker trips again, complete the following steps:
3. Disconnect the mainframe from the AC mains power source.
4. Remove all modules from the mainframe.
5. Complete the test procedure described in the Connecting to AC Mains
Power and Testing Power up section in Chapter 2, Installation and
Configuration.
6. If any voltages are outside the acceptable limits, contact National
Instruments.
7. If all voltages are within the acceptable limits, verify that your
PXI-1000 can meet the power requirements of your CompactPCI or
PXI modules. Overloading the chassis can cause the breaker to trip.
Refer to Appendix A, Specifications.
8. The over-current condition that caused the circuit breaker to trip may
be due to a faulty CompactPCI or PXI module. Refer to the
documentation that was supplied with the modules for troubleshooting
your modules.
© National Instruments Corporation
3-3
PXI-1000 User Manual
Chapter 3
Maintenance
Troubleshooting the PXI-1000
Refer to Table 3-1 to troubleshoot the PXI-1000 mainframe. The table lists
possible causes for power failure and recommends ways to correct the
problem.
Table 3-1. Troubleshooting
Possible Cause
What to Do
PXI-1000 mainframeis not Make sure that the PXI-1000 is connected
connected to power source. to a live electrical outlet. Try operating
another piece of equipment from this
outlet.
Power switches are not
switched on.
Set the rear (main) power switch and the
front (standby) power switch to the On
position.
Remote inhibiting input on Deactivate your system’s remote
the rear panel of the
mainframe is active.
inhibiting control.
Circuit breaker is tripped.
Reset the circuit breaker. Refer to the
Resetting the AC Mains Circuit Breaker
section in this chapter.
Power supply has failed.
Contact National Instruments.
PXI-1000 User Manual
3-4
© National Instruments Corporation
A
Specifications
This appendix contains complete specifications for the PXI-1000
mainframe.
Electrical
Table A-1. AC Input Specifications
Characteristic Description
Input Voltage Range
Input Frequency Range
Over-Current Protection
Operating Current
90–264 VAC
47 to 63 Hz
10 A circuit breaker
8 A
Line Regulation
± 0.5% over operating line range, all
outputs
Efficiency
80% typical
Transient Response
Output maximum excursion of ±5%
for 25% load step. Recovery less than
500 µs.
Power Disconnect
The rear main power switch supplies AC
power to power module. The front
(standby) power switch causes the power
module to supply DC power to the
CompactPCI/PXI backplane whenever
the main power switch (rear panel) is in
the On (1) position. The rear-panel D-sub
connector facilitates remote inhibiting
operation. Both the rear main power
switch and the front (standby) switch
must be in the On position prior to use of
remote inhibit. The power cord provides
main power disconnect.
© National Instruments Corporation
A-1
PXI-1000 User Manual
Appendix A
Specifications
Table A-2. DC Output Specifications
Characteristic Description
Maximum Usable Power
DC Current Capacity (IMP
300 W
Voltage
+3.3 V
+12 V
+5 V
)
IMP (Steady-State Current)
40 A
4 A
20 A
1 A
–12 V
Load Regulation
1% for 10% to 100% load changes,
all outputs
10% min. load required on V1 (+3.3 V)
for max. load regulation on outputs
V2 (+5 V), V3 (+12 V), and V4 (–12 V).
Maximum Ripple and
Noise
1% ripple, 1% noise, or 100 mV,
whichever is greater
20 MHz bandwidth
Over-Current Protection
Over-Voltage Protection
All outputs protected from short circuit
and overload, automatic recovery
3.3 V clamps at 20% to 30% above output
voltage
5 V, +12 V, and –12 V clamps at 10% to
20% above output voltage
Power Supply Unit MTBF 90,000+ hr
Power Supply/Fan Unit
MTTR
Replacement in under 5 minutes
PXI-1000 User Manual
A-2
© National Instruments Corporation
Appendix A
Specifications
Cooling
Table A-3. Cooling Specifications
Characteristic Description
Per Slot Cooling Capacity Slot cooling capacity in worst-case slot
is 20 W with fan speed set to HI
Module Cooling System
Forced air circulation (positive
pressurization) via two 60 cfm fans with
HI/LO speed selector
Slot Airflow Direction
P1 to P2, bottom of module to top of
module
Module Cooling Intake
Module Cooling Exhaust
Bottom rear of mainframe
Along both sides of mainframe
Forced air circulation via integrated fan
Power Supply Cooling
System
Power Supply Cooling
Intake
Rear of mainframe
Power Supply Cooling
Exhaust
Along both sides of mainframe rear and
upper rear panel
Module Cooling Fan
MTBF
40,000+ hr
Power Supply/Fan Unit
Replacement in under 5 minutes
Safety
Table A-4. Safety Specifications
Characteristic
Description
Safety Characteristics
UL 3111-1, IEC 1010-1, CSA 22.2
No. 1010.1
Installation Category II
Pollution Degree 2
Safety Class 1
© National Instruments Corporation
A-3
PXI-1000 User Manual
Appendix A
Specifications
Environmental
Table A-5. Environmental Specifications
Characteristic
Description
Operating Temperature
0° to 50° C (fan in LO position)
0° to 55° C (fan in HI position)
Storage Temperature
–20° to 70° C
Operating Relative
Humidity
Maximum 80% for temperatures up to
31° C, decreasing linearly to 50% at 40° C
Functional Shock
(Operating)
MIL-T-28800E CLASS 3, Half-Sine
Shock Pulse, 11 ms duration, 30 g peak
Operating Location
Indoor use
Random Vibration
(Operational)*
5 to 500 Hz, 0.31 gRMS
Random Vibration
(Non-Operational)*
10 to 500 Hz, 2.46 gRMS
EMC Emissions
EMC Immunity
Altitude
FCC Class A compliant and EN 55011
Group 1 Class A Compliant
Refer to DOC supplied with chassis for
compliance to relevant directives.
2 km (1.24 mi)
* Random vibration profiles were developed in accordance with MIL-T-28800E CLASS 3
and MIL-STD-810E Method 514 Test levels exceed those recommended in
MIL-STD-810E for Category 1 (Basic Transportation), Figures 514.4-1 through 514.4-3.
PXI-1000 User Manual
A-4
© National Instruments Corporation
Appendix A
Specifications
Backplane
Table A-6. Backplane Specifications
Characteristic Description
Size
3U-sized; one system slot (with three
system expansion slots) and seven
peripheral slots.
Compliant with IEEE 1101.10
mechanical packaging.
PXI Specification Revision 1.0
compliant.
Accepts both PXI and CompactPCI
(PICMG 2.0 R2.1) 3U modules.
Backplane Bare-Board
Material
UL 94 V-0 recognized
(File No. E 116551)
Backplane Connectors
Conform to IEC 917 and IEC 1076-4-101,
and are UL 94 V-0 rated
© National Instruments Corporation
A-5
PXI-1000 User Manual
Appendix A
Specifications
Mechanical
Table A-7. Mechanical Specifications
Characteristic Description
Overall Dimensions
Standard Mainframe
Height
Width
Depth
17.78 cm (7.00 in.)
27.02 cm (10.64 in.)
37.85 cm (14.90 in.)
Notes:
1.80 cm (.71 in.) added to height when
feet are installed.
When tilted with front feet extended on
table top, height is increased
approximately 5.29 cm (2.083 in.) in front
and 1.48 cm (.583 in.) in rear.
Weight
8.6 kg (19 lb.)
Maximum Module Weight 1.8 kg (4 lb.)
Materials
Sheet Aluminum (5052-H32) and
Cold Rolled Steel
Finish
Unpainted Aluminum
Conductive Clear Iridite
Clear Chromate Zinc Plating
Polyurethane Enamel
Cold Rolled Steel
Paint
PXI-1000 User Manual
A-6
© National Instruments Corporation
Appendix A
Specifications
Figure A-1 shows the PXI-1000 dimensions. The holes shown are for the
installation of the optional rack-mount kit. You can install this kit on the
front or rear of the chassis, depending on which end of the chassis you wish
to face toward the front of the instrument cabinet. Note that the front and
rear chassis mounting holes (size M4) are symmetrical.
ON
STANDBY
7.00 in.
2
3
4
5
6
7
8
1
10.64 in.
14.90 in.
1.823 in.
.78 in.
1.00 in.
3.541 in. 1.844 in.
.71 in.
Figure A-1. PXI-1000 Dimensions
© National Instruments Corporation
A-7
PXI-1000 User Manual
B
Pinouts
backplane.
slot.
Table B-2 shows the P2 (J2) connector pinout for the System Controller
slot.
Table B-3 shows the P1 (J1) connector pinout for the Star Trigger slot.
Table B-4 shows the P2 (J2) connector pinout for the Star Trigger slot.
Table B-5 shows the P1 (J1) connector pinout for the peripheral slots.
Table B-6 shows the P2 (J2) connector pinout for the peripheral slots.
Note
PXI signals are shown in boldface.
© National Instruments Corporation
B-1
PXI-1000 User Manual
Appendix B
Pinouts
Table B-1. P1 (J1) Connector Pinout for the System Controller Slot
Pin
25
24
23
22
21
20
19
18
17
16
15
12–14
11
10
9
Z
A
B
REQ64#
5V
C
ENUM#
V(I/O)
AD[3]
3.3V
D
E
F
GND
5V
3.3V
5V
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[1]
3.3V
AD[0]
5V
ACK64#
AD[2]
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[4]
GND
AD[7]
3.3V
AD[6]
M66EN
AD[11]
GND
AD[5]
AD[9]
GND
AD[8]
V(I/O)
AD[14]
3.3V
C/BE[0]#
AD[10]
AD[13]
C/BE[1]#
PERR#
LOCK#
TRDY#
AD[12]
3.3V
AD[15]
GND
SERR#
3.3V
PAR
SDONE
GND
SBO#
GND
DEVSEL#
3.3V
V(I/O)
IRDY#
Key Area
AD[16]
3.3V
STOP#
GND
FRAME#
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[18]
AD[21]
C/BE[3]#
AD[26]
AD[30]
REQ#
AD[17]
GND
GND
AD[20]
GND
AD[25]
GND
CLK
C/BE[2]#
AD[19]
AD[22]
AD[24]
AD[27]
AD[31]
GNT#
INTS
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
IDSEL
GND
AD[23]
V(I/O)
AD[28]
3.3V
8
7
AD[29]
GND
6
5
BRSVP1A5
BRSVP1A4
INTA#
BRSVP1B5
GND
RST#
GND
INTP
5V
4
V(I/O)
INTC#
TMS
3
INTB#
5V
INTD#
TDI
2
TCK
TDO
+12V
1
5V
–12V
TRST#
5V
PXI-1000 User Manual
B-2
© National Instruments Corporation
Appendix B
Pinouts
Table B-2. P2 (J2) Connector Pinout for the System Controller Slot
Pin
22
21
20
19
18
17
16
15
14
13
12
11
10
9
Z
A
PXI_RSVA22
RSV
B
PXI_RSVB22
GND
C
PXI_RSVC22
RSV
D
PXI_RSVD22
RSV
E
PXI_RSVE22
RSV
F
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
RSV
RSV
RSV
GND
RSV
RSV
GND
RSV
RSV
RSV
PXI_TRIG3
PXI_TRIG2
PXI_TRIG1
PXI_BRSVA15
AD[35]
PXI_TRIG4
GND
PXI_TRIG5
PRST#
DEG#
GND
PXI_TRIG6
GNT6#
PXI_TRIG7
GNT5#
AD[32]
AD[36]
AD[39]
AD[43]
AD[46]
AD[50]
AD[53]
AD[57]
AD[60]
PAR64
REQ6#
GND
PXI_TRIG0
GND
FAL#
REQ5#
GND
AD[34]
GND
AD[33]
V(I/O)
AD[38]
AD[37]
GND
AD[42]
AD[41]
GND
AD[40]
V(I/O)
AD[45]
AD[44]
GND
AD[49]
AD[48]
GND
AD[47]
V(I/O)
AD[52]
AD[51]
GND
8
AD[56]
AD[55]
GND
AD[54]
V(I/O)
7
AD[59]
AD[58]
GND
6
AD[63]
AD[62]
GND
AD[61]
V(I/O)
5
C/BE[5]#
V(I/O)
C/BE[4]#
GND
4
PXI_BRSVB4
GND
C/BE[7]#
GNT3#
SYSEN#
REQ1#
C/BE[6]#
GNT4#
REQ3#
REQ2#
3
CLK4
REQ4#
GNT2#
GNT1#
2
CLK2
CLK3
1
CLK1
GND
© National Instruments Corporation
B-3
PXI-1000 User Manual
Appendix B
Pinouts
Table B-3. P1 (J1) Connector Pinout for the Star Trigger Slot
Pin
25
24
23
22
21
20
19
18
17
16
15
12–14
11
10
9
Z
A
B
REQ64#
5V
C
ENUM#
V(I/O)
AD[3]
3.3V
D
E
F
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
5V
3.3V
AD[0]
5V
5V
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[1]
3.3V
AD[7]
3.3V
ACK64#
AD[2]
AD[4]
GND
AD[6]
AD[5]
AD[9]
GND
AD[8]
V(I/O)
AD[14]
3.3V
M66EN
AD[11]
GND
C/BE[0]#
AD[10]
AD[13]
C/BE[1]#
PERR#
LOCK#
TRDY#
AD[12]
3.3V
AD[15]
GND
SERR#
3.3V
PAR
SDONE
GND
SBO#
GND
DEVSEL#
3.3V
V(I/O)
IRDY#
STOP#
GND
FRAME#
Key Area
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[18]
AD[21]
C/BE[3]#
AD[26]
AD[30]
REQ#
AD[17]
GND
AD[16]
3.3V
GND
AD[20]
GND
AD[25]
GND
CLK
C/BE[2]#
AD[19]
AD[22]
AD[24]
AD[27]
AD[31]
GNT#
INTS
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
IDSEL
GND
AD[23]
V(I/O)
AD[28]
3.3V
8
7
AD[29]
GND
6
5
BRSVP1A5
BRSVP1A4
INTA#
BRSVP1B5
GND
RST#
GND
INTP
5V
4
V(I/O)
INTC#
TMS
3
INTB#
5V
INTD#
TDI
2
TCK
TDO
+12V
1
5V
–12V
TRST#
5V
PXI-1000 User Manual
B-4
© National Instruments Corporation
Appendix B
Pinouts
Table B-4. P2 (J2) Connector Pinout for the Star Trigger Slot
Pin
22
21
20
19
18
17
16
15
14
13
12
11
10
9
Z
A
B
PXI_RSVB22
GND
C
PXI_RSVC22
PXI_LBR1
PXI_STAR0
PXI_STAR3
PXI_TRIG5
PRST#
D
PXI_RSVD22
PXI_LBR2
GND
E
F
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
PXI_RSVA22
PXI_LBR0
PXI_LBR4
PXI_STAR2
PXI_TRIG3
PXI_TRIG2
PXI_TRIG1
PXI_BRSVA15
AD[35]
PXI_RSVE22
PXI_LBR3
PXI_STAR1
PXI_STAR5
PXI_TRIG6
PXI_CLK10
PXI_TRIG7
PXI_LBR6
AD[32]
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
PXI_LBR5
GND
PXI_STAR4
GND
PXI_TRIG4
GND
PXI_CLK10_IN
GND
PXI_TRIG0
GND
DEG#
FAL#
PXI_STAR6
GND
AD[34]
GND
AD[33]
AD[38]
V(I/O)
AD[37]
AD[36]
AD[42]
AD[41]
GND
AD[40]
GND
AD[39]
AD[45]
V(I/O)
AD[44]
AD[43]
AD[49]
AD[48]
GND
AD[47]
GND
AD[46]
AD[52]
V(I/O)
AD[51]
AD[50]
8
AD[56]
AD[55]
GND
AD[54]
GND
AD[53]
7
AD[59]
V(I/O)
AD[58]
AD[57]
6
AD[63]
AD[62]
GND
AD[61]
GND
AD[60]
5
C/BE[5]#
V(I/O)
V(I/O)
C/BE[4]#
GND
PAR64
4
PXI_BRSVB4
GND
C/BE[7]#
PXI_LBR8
SYSEN#
PXI_STAR10
C/BE[6]#
PXI_LBR10
PXI_STAR8
PXI_STAR12
3
PXI_LBR7
PXI_LBR11
PXI_STAR9
PXI_LBR9
PXI_STAR7
PXI_STAR11
2
PXI_LBR12
GND
1
© National Instruments Corporation
B-5
PXI-1000 User Manual
Appendix B
Pinouts
Table B-5. P1 (J1) Connector Pinout for the Peripheral Slot
Pin
25
24
23
22
21
20
19
18
17
16
15
12–14
11
10
9
Z
A
B
REQ64#
5V
C
ENUM#
V(I/O)
AD[3]
3.3V
D
E
F
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
5V
3.3V
AD[0]
5V
5V
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[1]
3.3V
AD[7]
3.3V
ACK64#
AD[2]
AD[4]
GND
AD[6]
AD[5]
AD[9]
GND
AD[8]
V(I/O)
AD[14]
3.3V
M66EN
AD[11]
GND
C/BE[0]#
AD[10]
AD[13]
C/BE[1]#
PERR#
LOCK#
TRDY#
AD[12]
3.3V
AD[15]
GND
SERR#
3.3V
PAR
SDONE
GND
SBO#
GND
DEVSEL#
3.3V
V(I/O)
IRDY#
STOP#
GND
FRAME#
Key Area
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
AD[18]
AD[21]
C/BE[3]#
AD[26]
AD[30]
REQ#
AD[17]
GND
AD[16]
3.3V
GND
AD[20]
GND
AD[25]
GND
CLK
C/BE[2]#
AD[19]
AD[22]
AD[24]
AD[27]
AD[31]
GNT#
INTS
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
IDSEL
GND
AD[23]
V(I/O)
AD[28]
3.3V
8
7
AD[29]
GND
6
5
BRSVP1A5
BRSVP1A4
INTA#
BRSVP1B5
GND
RST#
GND
INTP
5V
4
V(I/O)
INTC#
TMS
3
INTB#
5V
INTD#
TDI
2
TCK
TDO
+12V
1
5V
–12V
TRST#
5V
PXI-1000 User Manual
B-6
© National Instruments Corporation
Appendix B
Pinouts
Table B-6. P2 (J2) Connector Pinout for the Peripheral Slot
Pin
22
21
20
19
18
17
16
15
14
13
12
11
10
9
Z
A
B
PXI_RSVB22
GND
C
PXI_RSVC22
PXI_LBR1
PXI_LBL0
PXI_LBL3
PXI_TRIG5
PRST#
D
PXI_RSVD22
PXI_LBR2
GND
E
F
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
PXI_RSVA22
PXI_LBR0
PXI_LBR4
PXI_LBL2
PXI_TRIG3
PXI_TRIG2
PXI_TRIG1
PXI_BRSVA15
AD[35]
PXI_RSVE22
PXI_LBR3
PXI_LBL1
PXI_LBL5
PXI_TRIG6
PXI_CLK10
PXI_TRIG7
PXI_LBR6
AD[32]
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
PXI_LBR5
GND
PXI_LBL4
GND
PXI_TRIG4
GND
PXI_STAR
GND
PXI_TRIG0
GND
DEG#
FAL#
PXI_LBL6
GND
AD[34]
GND
AD[33]
AD[38]
V(I/O)
AD[37]
AD[36]
AD[42]
AD[41]
GND
AD[40]
GND
AD[39]
AD[45]
V(I/O)
AD[44]
AD[43]
AD[49]
AD[48]
GND
AD[47]
GND
AD[46]
AD[52]
V(I/O)
AD[51]
AD[50]
8
AD[56]
AD[55]
GND
AD[54]
GND
AD[53]
7
AD[59]
V(I/O)
AD[58]
AD[57]
6
AD[63]
AD[62]
GND
AD[61]
GND
AD[60]
5
C/BE[5]#
V(I/O)
V(I/O)
C/BE[4]#
GND
PAR64
4
PXI_BRSVB4
GND
C/BE[7]#
PXI_LBR8
SYSEN#
PXI_LBL10
C/BE[6]#
PXI_LBR10
PXI_LBL8
PXI_LBL12
3
PXI_LBR7
PXI_LBR11
PXI_LBL9
PXI_LBR9
PXI_LBL7
PXI_LBL11
2
PXI_LBR12
GND
1
© National Instruments Corporation
B-7
PXI-1000 User Manual
C
Customer Communication
For your convenience, this appendix contains forms to help you gather the information necessary
to help us solve your technical problems and a form you can use to comment on the product
documentation. When you contact us, we need the information on the Technical Support Form and
the configuration form, if your manual contains one, about your system configuration to answer your
questions as quickly as possible.
National Instruments has technical assistance through electronic, fax, and telephone systems to quickly
provide the information you need. Our electronic services include a bulletin board service, an FTP site,
a fax-on-demand system, and e-mail support. If you have a hardware or software problem, first try the
electronic support systems. If the information available on these systems does not answer your
questions, we offer fax and telephone support through our technical support centers, which are staffed
by applications engineers.
Electronic Services
Bulletin Board Support
National Instruments has BBS and FTP sites dedicated for 24-hour support with a collection of files
and documents to answer most common customer questions. From these sites, you can also download
the latest instrument drivers, updates, and example programs. For recorded instructions on how to use
the bulletin board and FTP services and for BBS automated information, call 512 795 6990. You can
access these services at:
United States: 512 794 5422
Up to 14,400 baud, 8 data bits, 1 stop bit, no parity
United Kingdom: 01635 551422
Up to 9,600 baud, 8 data bits, 1 stop bit, no parity
France: 01 48 65 15 59
Up to 9,600 baud, 8 data bits, 1 stop bit, no parity
FTP Support
To access our FTP site, log on to our Internet host, ftp.natinst.com, as anonymous and use
documents are located in the /support directories.
© National Instruments Corporation
C-1
PXI-1000 User Manual
Fax-on-Demand Support
Fax-on-Demand is a 24-hour information retrieval system containing a library of documents on a wide
range of technical information. You can access Fax-on-Demand from a touch-tone telephone at
512 418 1111.
E-Mail Support (Currently USA Only)
You can submit technical support questions to the applications engineering team through e-mail at the
Internet address listed below. Remember to include your name, address, and phone number so we can
contact you with solutions and suggestions.
Telephone and Fax Support
National Instruments has branch offices all over the world. Use the list below to find the technical
support number for your country. If there is no National Instruments office in your country, contact
the source from which you purchased your software to obtain support.
Country
Telephone
Fax
Australia
Austria
Belgium
Brazil
Canada (Ontario)
Canada (Québec)
Denmark
Finland
03 9879 5166
0662 45 79 90 0
02 757 00 20
011 288 3336
905 785 0085
514 694 8521
45 76 26 00
09 725 725 11
01 48 14 24 24
089 741 31 30
2645 3186
03 6120092
02 413091
03 5472 2970
02 596 7456
5 520 2635
03 9879 6277
0662 45 79 90 19
02 757 03 11
011 288 8528
905 785 0086
514 694 4399
45 76 26 02
09 725 725 55
01 48 14 24 14
089 714 60 35
2686 8505
France
Germany
Hong Kong
Israel
Italy
Japan
03 6120095
02 41309215
03 5472 2977
02 596 7455
5 520 3282
Korea
Mexico
Netherlands
Norway
Singapore
Spain
Sweden
Switzerland
Taiwan
0348 433466
32 84 84 00
2265886
91 640 0085
08 730 49 70
056 200 51 51
02 377 1200
01635 523545
512 795 8248
0348 430673
32 84 86 00
2265887
91 640 0533
08 730 43 70
056 200 51 55
02 737 4644
01635 523154
512 794 5678
United Kingdom
United States
PXI-1000 User Manual
C-2
© National Instruments Corporation
Technical Support Form
Photocopy this form and update it each time you make changes to your software or hardware, and use
the completed copy of this form as a reference for your current configuration. Completing this form
accurately before contacting National Instruments for technical support helps our applications
engineers answer your questions more efficiently.
If you are using any National Instruments hardware or software products related to this problem,
include the configuration forms from their user manuals. Include additional pages if necessary.
Name __________________________________________________________________________
Company _______________________________________________________________________
Address ________________________________________________________________________
_______________________________________________________________________________
Fax ( ___ ) ________________Phone ( ___ ) __________________________________________
Computer brand____________ Model ___________________Processor_____________________
Operating system (include version number) ____________________________________________
Clock speed ______MHz RAM _____MB
Display adapter __________________________
Mouse ___yes ___no Other adapters installed_______________________________________
Hard disk capacity _____MB Brand_________________________________________________
Instruments used _________________________________________________________________
_______________________________________________________________________________
National Instruments hardware product model _____________ Revision ____________________
Configuration ___________________________________________________________________
National Instruments software product ___________________ Version _____________________
Configuration ___________________________________________________________________
The problem is: __________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
List any error messages: ___________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
The following steps reproduce the problem: ___________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
PXI-1000 Hardware Configuration Form
Record the settings and revisions of your hardware and software on the line to the right of each item.
Complete a new copy of this form each time you revise your software or hardware configuration, and
use this form as a reference for your current configuration. Completing this form accurately before
contacting National Instruments for technical support helps our applications engineers answer your
questions more efficiently.
National Instruments Products
Complete Part Number ____________________________________________________________
Serial Number __________________________________________________________________
Line Voltage ____________________________________________________________________
Fan Speed ______________________________________________________________________
Safety Ground Connected? _________________________________________________________
Rack-Mount or Bench Top Configuration? ____________________________________________
Using Remote Power Monitoring ____________________________________________________
Filler Panels Installed in Empty Slots? ________________________________________________
Other Products
List and describe all devices installed in your mainframe.
Slot
Manufacturer, Description, and Function
1
2
3
4
5
6
7
8
Documentation Comment Form
National Instruments encourages you to comment on the documentation supplied with our products.
This information helps us provide quality products to meet your needs.
Title:
PXI™-1000 User Manual
Edition Date: January 1998
Part Number: 321710B-01
Please comment on the completeness, clarity, and organization of the manual.
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
If you find errors in the manual, please record the page numbers and describe the errors.
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
Thank you for your help.
Name _________________________________________________________________________
Title __________________________________________________________________________
Company _______________________________________________________________________
Address ________________________________________________________________________
_______________________________________________________________________________
E-Mail Address __________________________________________________________________
Phone ( ___ ) __________________________ Fax ( ___ ) _______________________________
Mail to: Technical Publications
National Instruments Corporation
Fax to: Technical Publications
National Instruments Corporation
512 794 5678
6504 Bridge Point Parkway
Austin, Texas 78730-5039
Glossary
Prefix
n-
Meanings
nano-
Value
10–9
10– 6
10–3
10–2
103
µ-
micro-
milli-
m-
c-
centi-
k-
kilo-
M-
mega-
106
Symbols
°
Degrees.
≥
≤
%
Equal or greater than.
Equal or less than.
Percent.
A
A
Amperes.
AC
Alternating current.
ANSI
AWG
American National Standards Institute.
American Wire Gauge.
B
backplane
An assembly, typically a printed circuit board, with connectors and signal
paths that bus the connector pins.
© National Instruments Corporation
G-1
PXI-1000 User Manual
Glossary
C
C
Celsius.
cfm
CFR
CSA
Cubic feet per minute.
Cooperative Fuel Research.
Canadian Standards Association.
D
daisy-chain
A method of propagating signals along a bus, in which the devices are
prioritized on the basis of their position on the bus.
DC
Direct current.
E
ECL
EIA
EMC
Emitter-coupled logic.
Electronic Industries Association.
Electromagnetic Compatibility.
F
FCC
Federal Communications Commission.
G
g
1) grams 2) A measure of acceleration equal to 9.8 m/s2.
General Purpose Interface Bus (IEEE 488).
GPIB
gRMS
A measure of random vibration. The root mean square of acceleration
levels in a random vibration test profile.
PXI-1000 User Manual
G-2
© National Instruments Corporation
Glossary
H
Hz
Hertz; cycles per second.
I
IEC
International Electrotechnical Commission; an organization that sets
international electrical and electronics standards.
IEEE
IMP
Institute of Electrical and Electronics Engineers.
Mainframe peak current.
Inches.
in.
L
lb
Pounds.
M
m
Meters.
MTBF
MTTR
Mean time between failure.
Mean time to repair.
N
NEMA
National Electrical Manufacturers Association.
PCI eXtensions for Instrumentation.
Relative humidity.
P
PXI
R
RH
© National Instruments Corporation
G-3
PXI-1000 User Manual
Glossary
RMS
Root mean square. A method used to measure electrical output in volts and
watts.
S
s
Seconds.
ST
Star Trigger.
Star Trigger slot
This slot is located at slot 2 and has a dedicated trigger line between each
peripheral slot. Use this slot for a module with ST functionality that can
provide individual triggers to all other peripherals.
System controller
A module configured for installation in Slot 0 of a VXIbus mainframe. This
device is unique in the VXIbus system in that it performs the VMEbus
system controller functions, including clock sourcing and arbitration for
data transfers across the backplane. Installing such a device into any other
slot can damage the device, the VXIbus backplane, or both.
U
UL
Underwriter’s Laboratories.
V
V
Volts.
VPP
VAC
Peak to peak voltage.
Volts alternating current.
W
W
Watts.
PXI-1000 User Manual
G-4
© National Instruments Corporation
Index
A
D
AC input (table), A-1
AC mains power
DB-9 connector
pinout (table), 2-5
connecting to, 2-3 to 2-4
resetting circuit breaker, 3-3
air cooling. See cooling.
power supply voltages (table), 2-4
DC output (table), A-2
documentation
conventions used in manual, xi
organization of manual, xi
related documentation, xii
B
bulletin board support, C-1
E
C
electrical specifications, A-1 to A-2
AC input (table), A-1
cables, power (table), 1-2
chassis initialization file, 2-7
circuit breaker, resetting, 3-3
cleaning
DC output (table), A-2
electronic support services, C-1 to C-2
e-mail support, C-2
exterior cleaning, 3-2
fan filters, 3-2
environmental specifications (table), A-4
equipment, optional, 1-2
interior cleaning, 3-2
precautions before cleaning, 3-1
CompactPCI
F
fan
installing modules (figure), 2-6
interoperability with PXI-1000 backplane,
1-3
cleaning fan filters, 3-2
setting speed, 2-2
configuration. See installation and configuration.
connector pinouts. See pinouts.
cooling
speed selector switch (figure), 1-5
fax and telephone support, C-2
Fax-on-Demand support, C-2
filler panel installation, 2-7
FTP support, C-1
air cooling of PXI-1000, 2-1 to 2-2
air intake (figure), 2-2
filler panel installation, 2-7
setting fan speed, 2-2
specifications (table), A-3
customer communication, xii, C-1 to C-2
G
ground, connecting, 2-3
© National Instruments Corporation
I-1
PXI-1000 User Manual
Index
precautions before cleaning, 3-1
preparation, 3-1
resetting AC mains circuit breaker, 3-3
service interval, 3-1
static discharge damage (caution), 3-1
troubleshooting, 3-4
I
Inhibit signal, turning off power supply
outputs, 2-5
installation and configuration, 2-1 to 2-7
chassis initialization file, 2-7
connecting safety ground, 2-3
connecting to AC mains power, 2-3 to 2-4
filler panel installation, 2-7
module installation, 2-6 to 2-7
CompactPCI or PXI module
(figure), 2-6
mechanical specifications, A-6 to A-7
O
optional equipment, 1-2
ejector/injector handle position
(figure), 2-7
rack mounting, 2-2
P
P1 (J1) connector pinouts
peripheral slot (table), B-6
Star Trigger slot (table), B-4
System Controller slot (table), B-2
P2 (J2) connector pinouts
peripheral slot (table), B-7
Star Trigger slot (table), B-5
System Controller slot (table), B-3
peripheral slots
remote power monitoring and inhibit
interface, 2-5
setting fan speed, 2-2
site considerations, 2-1 to 2-2
testing power up, 2-3 to 2-4
unpacking the PXI-1000, 1-1
J
overview, 1-5
J1 connector pinouts. See P1 (J1) connector
P1 (J1) connector pinouts (table), B-6
P2 (J2) connector pinouts (table), B-7
pinouts
pinouts.
J2 connector pinouts. See P2 (J2) connector
pinouts.
DB-9 connector (table), 2-5
P1 (J1) connector
peripheral slot (table), B-6
Star Trigger slot (table), B-4
System Controller slot (table), B-2
P2 (J2) connector
L
local bus
routing (figure), 1-6
signals, 1-6
peripheral slot (table), B-7
Star Trigger slot (table), B-5
System Controller slot (table), B-3
power cables (table), 1-1 to 1-2
power monitoring connector. See DB-9
connector.
M
maintenance of PXI-1000, 3-1 to 3-4
cleaning
exterior cleaning, 3-2
fan filters, 3-2
power problems, troubleshooting, 3-4
interior cleaning, 3-2
PXI-1000 User Manual
I-2
© National Instruments Corporation
Index
power supply
remote power monitoring and inhibit
S
safety ground, connecting, 2-3
safety specifications (table), A-3
service interval, 3-1
specifications, A-1 to A-7
backplane, A-5
interface, 2-5
universal, 2-3
voltages at power monitoring connector
(DB-9) (table), 2-4
power up, testing, 2-3 to 2-4
PXI-1000
cooling, A-3
electrical, A-1 to A-2
environmental, A-4
mechanical, A-7
cooling air intake (figure), 2-2
dimensions (figure), A-7
features, 1-2 to 1-3
front view (figure), 1-4
installation (See installation and
configuration)
safety, A-3
Star Trigger (ST) slot
description, 1-5
P1 (J1) connector pinouts (table), B-4
P2 (J2) connector pinouts (table), B-5
triggers for peripheral slots (figure), 1-6
static discharge damage (caution), 3-1
System Controller slot
maintenance (See maintenance of
PXI-1000)
optional equipment, 1-2
power cables required (table), 1-1 to 1-2
rear view (figure), 1-5
requirements for getting started,
1-1 to 1-2
description, 1-5
P1 (J1) connector pinouts (table), B-2
P2 (J2) connector pinouts (table), B-3
system reference clock, 1-7
specifications (See specifications)
unpacking, 1-1
PXI-1000 backplane, 1-3 to 1-7
interoperability with CompactPCI, 1-3
local bus, 1-6
T
technical support, C-1 to C-2
telephone and fax support, C-2
trigger bus, 1-6 to 1-7
peripheral slots, 1-5
specifications (table), A-5
Start Trigger slot, 1-5
troubleshooting the PXI-1000, 3-4
System Controller slot, 1-5
system reference clock, 1-7
PXI_CLK10, 1-7
U
PXI_CLK10_IN pin, 1-7
unpacking the PXI-1000, 1-1
R
V
rack mounting, 2-2
voltages at power monitoring connector
(DB-9) (table), 2-4
remote power monitoring and inhibit
interface, 2-5
© National Instruments Corporation
I-3
PXI-1000 User Manual
|